7. Detailed Description

Accurate Setting of Temperature Limit

If a more accurate temperature limit is required, for example 50 °C, follow the procedure below.

1.Set the temperature limit switch to its highest value.

2.Set the working temperature to 50 °C according to Section 4.4.

3.Run the unit and wait until it reaches exactly 50 °C.

4.Turn the temperature limit dial very slowly anticlock- wise (temperature down) until the safety circuit is trigge- red and the display indicates a malfunction.

5.Change the working temperature setting to any value lower than 50 °C.

6.Wait until the bath liquid has cooled down to below 50 °C.

7.Press the reset key.

The circulator can now be used for temperatures up to

50 °C. As soon as 50 °C is reached, the unit switches off.

Protecting the pump motor

When the temperature in the winding of the pump motor reaches a certain level, the general cut-off will be activated. This protection prevents the motor from overheating. The following error message will be displayed.

. After the winding of the pump has cooled

down, the following code

is displayed.

Check the pump and wait a few minutes until the motor tem- perature has dropped further before restarting the unit.

If this fault happens frequently, contact service personnel.

Sensor check

The temperature sensor is permanently monitored. The gene- ral cut-off is activated if sensor breakage or current leakage is detected. This fault is

indicated by

.