4.4Tempera- ture setting

4. Operation

4.4Tempera- ture setting

4.5Tempera- ture correction value

2.Use Manual background or Manual background to set the maximum allowed set tempe- rature. The factory pre-set value is 90.0 °C.

3.Confirm with Manual background before the display returns to the actu- al temperature value.

Note: The new value is not saved until

has been pres-

sed.

 

1. Press Manual background to select the S mode. Manual backgroundManual background

2.Use Manual background or Manual background to set the working temperature.

3.Confirm with Manual background before the display automatically returns to the actual temperature value.

Note: The new set value is not saved until

has been

pressed.

 

The display continuously indicates the actual temperature in the bath at the position of the sensor. This temperature does not correspond directly to the temperature in the externally connected equipment. Under most circumstances this tempe- rature difference is very small. To adjust the display value to reflect the temperature in the external system, proceed as fol- lows:

1.Measure the temperature in the external system.

2.Calculate the temperature difference between this value and the value displayed.

Tdiff = Tdisp -TmeasTdiff is the correction value “c” that has to be entered.

3. Press Manual background to select the C mode

4.Use Manual background or Manual background to set the correction value with its + or - sign. The range is -25.0 to +2.50 °C. The factory pre-set value is 0.0.

11

Page 13
Image 13
TRENDnet 18-1106-33 user manual Tempera- ture setting 4.5 Tempera- ture correction value