4. Operation

4.4Tempera- ture setting

4.5Tempera- ture correction value

2.Use or to set the maximum allowed set tempe- rature. The factory pre-set value is 90.0 °C.

3.Confirm with before the display returns to the actu- al temperature value.

Note: The new value is not saved until

has been pres-

sed.

 

1. Press to select the S mode.

2.Use or to set the working temperature.

3.Confirm with before the display automatically returns to the actual temperature value.

Note: The new set value is not saved until

has been

pressed.

 

The display continuously indicates the actual temperature in the bath at the position of the sensor. This temperature does not correspond directly to the temperature in the externally connected equipment. Under most circumstances this tempe- rature difference is very small. To adjust the display value to reflect the temperature in the external system, proceed as fol- lows:

1.Measure the temperature in the external system.

2.Calculate the temperature difference between this value and the value displayed.

Tdiff = Tdisp -TmeasTdiff is the correction value “c” that has to be entered.

3. Press to select the C mode

4.Use or to set the correction value with its + or - sign. The range is -25.0 to +2.50 °C. The factory pre-set value is 0.0.

11

Page 13