Compaq mar345 manual Min, Max

Page 31

mar345 Program Manual

3 1

______________________________________________________________________________________

The functions of the mar345−Colors window widgets are:

Grey scales Selection of coloring mode. Alternatives: Blue scales, Rainbow.

Min

All pixels with intensities <= Min are drawn white (black in Blue scales).

 

The dashed line in the histogram plot moves to the specified value.

Max

All pixels with intensities >= Max are drawn black (Grey scales), white

 

(Blue scales) or red (Rainbow). Intensities which are above the saturation

 

limit (128000) are drawn in green. The dashed line in the histogram plot

 

moves to the specified value.

In the histogram plot, the mouse buttons have the following functions:

Left mouse button press Places the dashed red line in the histogram plot to a new position marking the minimum. The value in the "Min" text field is changed accordingly.

Right mouse button press Places the dashed red line in the histogram plot to a new position marking the maximum. The value in the "Max" text field is changed accordingly.

Center mouse button drag Changes color tables. Try it out to see what it really does. This works for grey and blue scales only and for 8−bit color displays only!

Note:

By changing Min or Max or setting these values with the mouse button the color distribution for the whole image will be recalculated. This takes some CPU−time, so please don’t d o more than one change at a time.

Image 31
Contents Mar research All Rights Reserved Mar345 Program Manual Version February 24Contents Computer Requirements IntroductionEnvironment Running mar345 Command LineOutput Files Input FilesMar.lp Basic Concepts and RulesMar.log Mar.spyStarting Up Main Window User InterfaceMenu Bar Windows Submenu Menu BarScanner Status Area Scanner Command Area Editing Data Collection Parameters Directory Scan modeCollect mode ItemDescription TitleOscillations ItemDescription First image noNo. of images Exp. timeWavelength variable Loading and Saving Data Collection Parameters Moving Distance and Phi Stopping Data Collection Starting Data CollectionSingle Scan Shutter TimerHardware Status Getting Help14 X−ray Setup Errors and Warnings Display Window Image DisplayShow/Hide statistics Options SubmenuFiles ColorsNext image Reset colorsDo not/Keep color scales Turn On/Off 3D−plotImage Area Mar345 Program Manual Zoom Buttons Work AreaInformation Areas Load ButtonsRainbow Window TitleColors Window Blue scalesMax MinNo. of peaks Cross−Section WindowMessage area Lower horizontal axis Moved using the right mouse buttonVertical axis Upper horizontal axisStrategies Data CollectionImageroot.html Checking ParametersEnd of Data Collection Imageroot.SUMMARYReason TroubleshootingNo. Description MessageTabel from previous 1101 Appendix Configuration File Calibration Files