Tabel continued from previous page.

_______________________________________________________________________

No. Description

_______________________________________________________________________

1101

Message:

"No scan modes found in nb_code ..."

1102

 

"Something wrong with byteorder in nb_code ..."

1105

Reason:

"No suitable scanning mode found in nb_code".

 

File $MARTABLEDIR/mar2300.XXX and/or

 

Action:

$MARTABLEDIR/mar3450.XXX may be corrupted or empty.

 

Check file sizes (73 MB and 103MB). Use command:

 

 

catmar $MARTABLEDIR/mar2300.XXX

 

 

to look at the calibration file header.

1103

Message:

"Scanner serial number in nb_code differs from config"

 

Reason:

The scanner no. in the file headers of files

 

 

$MARTABLEDIR/mar2300.XXX and/or

 

 

$MARTABLEDIR/mar3450.XXX are not identical to

 

 

$MAR_SCANNER_NO. The calibration files may not

 

Action:

belong to the scanner.

 

Call.

1110

Message:

"Cannot create image file"

1111

 

"Cannot open image file"

1112

 

"Error writing image array"

1115

Reason:

"Error writing image header"

 

Insufficient disk space or invalid file permissions.

 

Action:

Clean up disk or check data directory permissions.

1215/1241

Message:

"Cannot OPEN SHUTTER at START of EXPOSURE"

1234

 

"SHUTTER was already OPEN at START of EXPOSURE!"

1238

Reason:

"Cannot CLOSE SHUTTER at END of EXPOSURE!"

 

X−ray beam shutter damaged or dirty. Otherwise, the manual

 

 

shutter operation switch mounted on the base may have

 

Action:

been left in the "Open" state.

 

Check shutter and manual shutter switch on the base.

 

 

Open and close the shutter manually. Is it very hard

 

 

to operate?

1239

Message:

"STEPPER error during EXPOSURE!"

 

Reason:

Problem with PHI axis movement. Possibly software related.

 

Action:

Mail mar.spy file. Call.

_______________________________________________________________________

Image 38
Contents Mar research Mar345 Program Manual Version February 24 All Rights ReservedContents Environment IntroductionComputer Requirements Command Line Running mar345Input Files Output FilesMar.spy Basic Concepts and RulesMar.log Mar.lpStarting Up Menu Bar User InterfaceMain Window Menu Bar Windows SubmenuScanner Status Area Scanner Command Area Editing Data Collection Parameters ItemDescription Title Scan modeCollect mode DirectoryExp. time ItemDescription First image noNo. of images OscillationsWavelength variable Loading and Saving Data Collection Parameters Moving Distance and Phi Starting Data Collection Stopping Data CollectionShutter Timer Single ScanGetting Help Hardware Status14 X−ray Setup Errors and Warnings Image Display Display WindowColors Options SubmenuFiles Show/Hide statisticsTurn On/Off 3D−plot Reset colorsDo not/Keep color scales Next imageImage Area Mar345 Program Manual Load Buttons Work AreaInformation Areas Zoom ButtonsBlue scales Window TitleColors Window RainbowMin MaxMessage area Cross−Section WindowNo. of peaks Upper horizontal axis Moved using the right mouse buttonVertical axis Lower horizontal axisData Collection StrategiesImageroot.SUMMARY Checking ParametersEnd of Data Collection Imageroot.htmlMessage TroubleshootingNo. Description ReasonTabel from previous 1101 Appendix Configuration File Calibration Files