5. Data Collection

5.1 Strategies

While some general rules apply how to collect data best, a sensible choice of data collection parameters depends on individual circumstances, i.e. crystal quality, beam properties, etc. The following section gives some hints on how to program a data collection:

Distance crystal−detector:

The distance crystal−detector must be chosen such that spots don’t overlap. It should be chosen such that the highest resolution of the spots is close to the edge of the scanned area, i.e. move the detector back as far as possible. This gives a better Intensity/σ ratio!

ϕ per image:

The ∆ ϕ must be chosen such that spots don’t overlap. If the background is large compared to the average spot intensity (i.e. most protein crystals) choose relatively small values (0.5 to

2.0deg.). The smaller, the better the integration works, but the more exposures you need to get a complete data set.

Number of images:

How many images you need depends on the symmetry of the space group and on the setting of the crystal. As a rule of thumb, orthorhombic crystals require 90.0 deg. of data, monoclinic ones 120 deg., triclinic ones 180 deg., trigonal ones 60.0 deg., tetragonal ones 45.0 deg, hexagonal ones about 30.0 deg. and cubic ones about 25.0 deg.. High resolution data collections usually require a wedge of data − say 20 deg. − at another crystal setting becaus e of the blind region.

Exposure time:

The exposure time should be less than 30 minutes, in any case, because of partial decay of the F−centers on the imaging plate. For very weak diffractors which require longer exposur e times it is better to scan one exposure after 30 min. and to repeat the same exposure and add the images together later. Otherwise, exposure times should be chosen such that the dynamic range (0:130000) is not exceeded. It is actually best to have maximum intensities considerably less, e.g. 60000.

Number of oscillations:

Best to leave at 1, except when using exposure times larger than 15 minutes.

Scan mode:

Why scan the 345 mm plate if there is not enough diffraction? A 180 mm scan takes half the time to scan, needs 80% less disk space and 80% less processing time!

Image 34
Contents Mar research Mar345 Program Manual Version February 24 All Rights ReservedContents Computer Requirements IntroductionEnvironment Command Line Running mar345Input Files Output FilesMar.spy Basic Concepts and RulesMar.log Mar.lpStarting Up Main Window User InterfaceMenu Bar Menu Bar Windows SubmenuScanner Status Area Scanner Command Area Editing Data Collection Parameters ItemDescription Title Scan modeCollect mode DirectoryExp. time ItemDescription First image noNo. of images OscillationsWavelength variable Loading and Saving Data Collection Parameters Moving Distance and Phi Starting Data Collection Stopping Data CollectionShutter Timer Single ScanGetting Help Hardware Status14 X−ray Setup Errors and Warnings Image Display Display WindowColors Options SubmenuFiles Show/Hide statisticsTurn On/Off 3D−plot Reset colorsDo not/Keep color scales Next imageImage Area Mar345 Program Manual Load Buttons Work AreaInformation Areas Zoom ButtonsBlue scales Window TitleColors Window RainbowMin MaxNo. of peaks Cross−Section WindowMessage area Upper horizontal axis Moved using the right mouse buttonVertical axis Lower horizontal axisData Collection StrategiesImageroot.SUMMARY Checking ParametersEnd of Data Collection Imageroot.htmlMessage TroubleshootingNo. Description ReasonTabel from previous 1101 Appendix Configuration File Calibration Files