mar345 Program Manual | 3 7 |
______________________________________________________________________________________
Tabel continued from previous page.
_______________________________________________________________________
No. Description
_______________________________________________________________________
1050 | Message: | "SHUTTER did not work properly. Abandoning data collection" |
| Reason: | X−ray beam shutter damaged or dirty. |
| Action: | Check shutter. Open and close it manually. Is it very hard |
|
| to operate? |
1060 | Message: | "X−ray reading too low ..." |
| Reason: | The X−ray intensity as read from the ion. chamber is |
|
| below the critical level (see configuration file: |
|
| INTENSITY MIN xxx). The data collection starts only |
| Action: | if the X−ray reading is above the configured value. |
| Check generator, monochromator or collimator. | |
1061 | Message: | "X−rays reading too high for DOSE mode ..." |
| Reason: | The X−ray intensity as read from the ion. chamber shows |
|
| very large numbers (>100). The work in DOSE mode becomes |
| Action: | inaccurate. |
| Turn down gain selector of ion. chambers. | |
1070 | Message: | "Could not recover from previous errors after X retries" |
| Reason: | Follow up of previous messages. |
| Action: | Check previous messages. |
1080 | Message: | "SHUTTER did not open during exposure" or |
| Reason: | "Shutter did not close at end of exposure" |
| X−ray beam shutter damaged or dirty. | |
| Action: | Check shutter. Open and close it manually. Is it very hard |
|
| to operate? |
1090 | Message: | "Scanner home position (X) differs from Y +/− dy" |
| Reason: | Problem with radial movement of reading head. |
| Action: | Mail mar.spy. Call. |
1100 | Message: | "Cannot open nb_code!" |
| Reason: | File $MARTABLEDIR/mar2300.XXX and/or |
|
| $MARTABLEDIR/mar3450.XXX are missing or don’t |
| Action: | have read permission. |
| Check existence of files and permissions. |
_______________________________________________________________________