Compaq mar345 manual Tabel from previous

Page 37

mar345 Program Manual

3 7

______________________________________________________________________________________

Tabel continued from previous page.

_______________________________________________________________________

No. Description

_______________________________________________________________________

1050

Message:

"SHUTTER did not work properly. Abandoning data collection"

 

Reason:

X−ray beam shutter damaged or dirty.

 

Action:

Check shutter. Open and close it manually. Is it very hard

 

 

to operate?

1060

Message:

"X−ray reading too low ..."

 

Reason:

The X−ray intensity as read from the ion. chamber is

 

 

below the critical level (see configuration file:

 

 

INTENSITY MIN xxx). The data collection starts only

 

Action:

if the X−ray reading is above the configured value.

 

Check generator, monochromator or collimator.

1061

Message:

"X−rays reading too high for DOSE mode ..."

 

Reason:

The X−ray intensity as read from the ion. chamber shows

 

 

very large numbers (>100). The work in DOSE mode becomes

 

Action:

inaccurate.

 

Turn down gain selector of ion. chambers.

1070

Message:

"Could not recover from previous errors after X retries"

 

Reason:

Follow up of previous messages.

 

Action:

Check previous messages.

1080

Message:

"SHUTTER did not open during exposure" or

 

Reason:

"Shutter did not close at end of exposure"

 

X−ray beam shutter damaged or dirty.

 

Action:

Check shutter. Open and close it manually. Is it very hard

 

 

to operate?

1090

Message:

"Scanner home position (X) differs from Y +/− dy"

 

Reason:

Problem with radial movement of reading head.

 

Action:

Mail mar.spy. Call.

1100

Message:

"Cannot open nb_code!"

 

Reason:

File $MARTABLEDIR/mar2300.XXX and/or

 

 

$MARTABLEDIR/mar3450.XXX are missing or don’t

 

Action:

have read permission.

 

Check existence of files and permissions.

_______________________________________________________________________

Image 37
Contents Mar research All Rights Reserved Mar345 Program Manual Version February 24Contents Computer Requirements IntroductionEnvironment Running mar345 Command LineOutput Files Input FilesMar.log Basic Concepts and RulesMar.spy Mar.lpStarting Up Main Window User InterfaceMenu Bar Windows Submenu Menu BarScanner Status Area Scanner Command Area Editing Data Collection Parameters Collect mode Scan modeItemDescription Title DirectoryNo. of images ItemDescription First image noExp. time OscillationsWavelength variable Loading and Saving Data Collection Parameters Moving Distance and Phi Stopping Data Collection Starting Data CollectionSingle Scan Shutter TimerHardware Status Getting Help14 X−ray Setup Errors and Warnings Display Window Image DisplayFiles Options SubmenuColors Show/Hide statisticsDo not/Keep color scales Reset colorsTurn On/Off 3D−plot Next imageImage Area Mar345 Program Manual Information Areas Work AreaLoad Buttons Zoom ButtonsColors Window Window TitleBlue scales RainbowMax MinNo. of peaks Cross−Section WindowMessage area Vertical axis Moved using the right mouse buttonUpper horizontal axis Lower horizontal axisStrategies Data CollectionEnd of Data Collection Checking ParametersImageroot.SUMMARY Imageroot.htmlNo. Description TroubleshootingMessage ReasonTabel from previous 1101 Appendix Configuration File Calibration Files