MTEK6000 SERIES USER'S MANUAL
NOTE
When you have changed the input drive The index box assembly contains a reed value, be sure tocorrespondingmove the existing drive
switch (G) and amagnet (F). value label from the window and replace it The magnet should be positioned so there is with a new label that states the current input 0.07" - 0.1" clearance between the magnet and drive value. For your convenience, switch. To adjust this clearance, refer to Fig. Metretek, Inc. supplies extra labels with the
Magnet Sensor Adjustment
The index box assembly contains a reed switch
(G)and a corresponding magnet (F). The magnet should be positioned so there is
Figure 2- 3 Base Plate Showing Unit & Index Rotation
UNIT AND INDEX ROTATION
In general, the label side of the index base plate (front) should face the front of the meter. This allows the MTEK6000 EFC to also face the front of the meter. In certain applications, the MTEK6000 EFC and index can be installed 1800 from the standard position so that viewing of the MTEK6000 EFC is acceptable. To rotate the unit and index, refer to Fig.
1. Remove the 4 bolts at location A.
1.Loosen set screw h.
2.Move the magnet until clearance is correct.
3.Retighten set screw h.
4. Verify the clearance is still correct. If not, repeat the procedure.
2.Rotate the base plate 1800 clockwise so that the front label side is now facing the rear of the unit.
3.Insert the 4 bolts into location B with the two lockwashers as shown. Fig.
COUNTER MASKING
Figure 2- 4 Uncorrected Counter Masking for Eight Digit Counter
January 2002 |