Compaq MTEK6000 user manual Cubic Unit/Pulse In or Meter Drive, Date Month, Day, Year

Page 61

MTEK6000 SERIES USER'S MANUAL

Cubic Unit/Pulse In or Meter Drive

This parameter determines the volume unit represented by one input pulse, and is normally set to equal the drive (CF/Rev) of the meter. Standard indexes produce one pulse per revolution; therefore, the CF/Pulse will equal the drive rate of the meter. The Default value is 100.

For instruments utilizing remote pulsing devices, this value can also be determined by dividing the CF/Rev of the meter by the number of pulses per revolution generated by the remote pulser.

For example: Consider a 3GT meter with a 100 CF/Rev drive and a 50 Pulse/Rev Imac Pulsamatic Transmitter. The CF/Pulse is determined as follows:

CF/Pulse = 100 CF/Rev * 1 Rev/50 Pulses = 2 CF/Pulse

NOTE FOR METERS WITH 5 CF/REV DRIVE RATES:

Re-position the input compound gear to the correct position, and set this parameter to a value of 10. Refer to Section Two or contact Metretek, Inc. for assistance in positioning the compound gear.

Current Day High Volume Alarm Reset After the Current Day's Total is exceeded, the setpoint at which the unit exits this condition is entered in this location. The Default value is 99,990.

Current Day High Volume Alarm Setpoint

This parameter specifies the setpoint at which the unit determines whether the Current Day Flow Total is in an alarm condition. For transport or interruptible customers, this parameter can be used to alarm when an account has exceeded a predetermined daily volume allocation. The Default value is 100,000.

January 2002

Date (Month, Day, Year)

This is the current Date in the unit (MMDDYY). It is updated on each process scan.

Flow Rate Update Interval (s)

This parameter determines how often the flow rate gets updated in the EFC. However, it has no control over Volume updates. Volume is updated every time the unit wakes up from the sleep mode or every 3-4 seconds if the unit is always awake.

The Flow Rate Update Interval parameter is of the form XXX.YY.M where the integer portion (XXX) is the time interval in seconds between Flow Rate updates. The digits after the decimal point (YY.M) represent the zero- flow period (i.e. the amount of time to wait without receiving pulses before generating a zero flow rate). The first two digits are minutes and the third digit (T) is a fraction of

aminute. Thus, a two-minute zero-flow period would be entered as XXX.020, a ten minute period as XXX.10. The maximum allowed period is 27 minutes and the default is one minute (XXX.010).

The user MUST specify a zero-flow period, which is larger than or equal to the Flow Rate Update Interval. Specifying a zero-flow period smaller than the Flow Rate Update Interval will result in a zero-flow rate being displayed even when there is an apparent flow. For example: A value of 600.11 would update flow rate every 10 minutes and display zero flow if 11 minutes passed without an input pulse. A value of 600.01 will not work properly. It would try to update flow rate every 10 minutes, but would zero the flow rate every minute, even if pulses come in faster than one per minute.

C - 3

Image 61
Contents MTEK6000 Series Table of Contents Optional Equipment List of Figures January January Hazardous Locations OverviewCompliance ONE-YEAR WarrantySecurity Options MTEK6000 exterior view MTEK6000 interior view Initial Check OUT UnpackingOptional Power Supplies Power for the MTEK6000Mounting the Instrument on the Meter Output Shaft Rotation Setting UP the Index AssemblyCounter Masking Unit and Index RotationWall and Pipe Mounting Pulse Input to the MTEK6000Wall Mounting Transducers in the MTEK6000 Connecting the Pressure TubingTypical Installation for MTEK6000 EFC Pipe Size Insertion Thermowell Length Part Number Installing the Thermal ProbePulse Output Wiring Installing the Pulse Output WiringUncorrected Pulse Output Specifications Volume and Alarm Pulse SpecificationsCommunications Connecting the Telephone LineRS-232C Serial Direct Communications 9600 Baud Modem Communications 2400 BaudActivity Indicator MTEK6000 Function RS232 cable connected January References GroundingJanuary Display Mode SLEEP/WAKE-UP ModeViewing and clearing alarms using the magnet Alarm ModeFirst Time Power Alarm Faulty Counter Alarm EFCHigh Flow Rate Alarm EFC & EFM High Differential Pressure Alarm EFMLow Supply Volts Alarms Low Pressure AlarmLow Temperature Alarm Lost Temperature AlarmSoftware Error Alarm Open Door Alarm optionalMemory History Logging Editing Parameters Configuration ModeOptional Keypad and Display Audit Trail Assigning Function KeysWake-Up on Pulse Event Driven Analog SamplingSpecial Key Combinations Calibrating the Pressure Transducer Calibration ModeZero NEW? Calibrating the Temperature TransducerCalibrating the Differential Pressure Transmitter EFM Only Analog Output AO Option Optional EquipmentAnalog Output Option MTEK6000 Analog Output Specifications Zero x.xxx 04.000 +y.yyy mA Changing the Battery Enclosure MaintenanceCalibration PcGas Customer Monitor PcGas Meter Utility PackageMtek Manager DC2000 January Label I.D Description Address Alarms Alarm CodeTable A-5 Standard alarms for MTEK6000 EFC w/ Aux Pressure Label Description Address Aux Pressure Table A-11 Standard alarms for MTEK6000 EFC2 w/2 Aux Press Table A-14 Standard alarms for MTEK6000 EFM w/ Aux Press MTEK6000 EPR MTEK6000 ETR MTEK6000 Eptr January NX19 Supercompressibility Report AGA-7 Volume CalculationsJanuary Alarm Pulse Time ms Alarm Pulse Output Enable/DisableAudit Trail Logging Enable Analog Sampling Rate secondsCalibrate Mode Time-out Base TemperatureCalibration Password Configuration PasswordCubic Unit/Pulse In or Meter Drive Flow Rate Update Interval sDate Month, Day, Year High Pressure Alarm Setpoint High Flow Rate Alarm SetpointHigh Pressure Alarm Reset High Temperature Alarm ResetLow Supply Voltage Alarm Reset Low Pressure Alarm SetpointLow Supply Voltage Alarm Setpoint Low Temperature Alarm ResetTime Hhmmss Hours, Minutes, Seconds Uncorrected VolumeWake Up Interval seconds Wake Up On Pulse event driven EFCTemperature Differential Flow & VolumeUnits of Measure AccumulationsJanuary SBC Revision a Corrector Board Description Jumper position OPT Revision a Option BoardPulse Input #1 PLI Revision C Phone Line Interface Consumer Information and FCC Requirements Service RequirementWarranty & Repair Service EL0001 Sheet 1 of 2 Class I Division 2 Installation EL0001 Sheet 2 of 2 Class I Division 2 Installation Warranty Information January Appendix G TII Station Protector January Appendix H Hazardous Area Installation Control Drawings October October