Compaq MTEK6000 user manual Calibrating the Differential Pressure Transmitter EFM Only

Page 38

MTEK6000 SERIES USER'S MANUAL

Calibrating the Differential Pressure Transmitter (EFM Only)

NOTE

Pressing esc repeatedly from anywhere within the calibration procedure will back the operator out of calibration mode.

1.Display the differential pressure by pressing F8 or jumping to 040302.

2.Press cal. Enter your password at the optional PASSWORD? prompt, if required.

3.The unit will enter calibration mode. The display will alternate between CALIBRATING and the parameter label (Diff Press “H20 for example).

4.Open both pressure connection valves.

5.Open bypass valve and close both high pressure and low-pressure transmitter connection valves on manifold.

6.Slowly open the high-pressure transmitter connection valve and allow the transmitter’s output to stabilize.

7.Press zero. The display now shows:

ZERO> XX.XXX

NEW?>

XX.XXX represents the unit's default zero value. If the current zero reference matches the unit's default, simply press ent to collect the new point. Otherwise, key in the value of the current reference before pressing ent. The unit should display Calculating . . .

briefly, and then display the new point. Pressing esc instead of ent at this point aborts the operation and returns you to the calibration prompt.

8. Apply the span (full scale) reference to

the differential pressure sensor and wait for the reading to stabilize

9. Press span. The unit now shows:

SPAN> XX.XXX

NEW?>

As with the zero point, if the external reference matches the default span value, simply press ent. Otherwise, key in the current value of the external reference, then press ent. After pressing ent, the display reading should immediately adjust to reflect the new calibration point. Pressing esc instead of ent at this point aborts the operation and returns the operator to the calibration prompt.

10.Steps 5 - 9 are required only once. They may be repeated as often as necessary while in calibration, but only the most recent point will be saved on completion of calibration.

11.To permanently store the results of the calibration press ent, and the unit will prompt Enter to accept Calibration. Simply press ent again to save the calibration. Press esc to abort the calibration.

12.Confirm that the bypass valve is open.

13.Slowly open the high-pressure valve.

14.Open the low-pressure valve.

15.Close the bypass valve.

3-12

January 2002

Image 38
Contents MTEK6000 Series Table of Contents Optional Equipment List of Figures January January Overview Hazardous LocationsSecurity Options ONE-YEAR WarrantyCompliance MTEK6000 exterior view MTEK6000 interior view Unpacking Initial Check OUTMounting the Instrument on the Meter Power for the MTEK6000Optional Power Supplies Setting UP the Index Assembly Output Shaft RotationUnit and Index Rotation Counter MaskingPulse Input to the MTEK6000 Wall and Pipe MountingWall Mounting Connecting the Pressure Tubing Transducers in the MTEK6000Typical Installation for MTEK6000 EFC Installing the Thermal Probe Pipe Size Insertion Thermowell Length Part NumberInstalling the Pulse Output Wiring Pulse Output WiringVolume and Alarm Pulse Specifications Uncorrected Pulse Output SpecificationsRS-232C Serial Direct Communications 9600 Baud Connecting the Telephone LineCommunications Modem Communications 2400 BaudActivity Indicator MTEK6000 Function RS232 cable connected January Grounding ReferencesJanuary SLEEP/WAKE-UP Mode Display ModeAlarm Mode Viewing and clearing alarms using the magnetHigh Flow Rate Alarm EFC & EFM Faulty Counter Alarm EFCFirst Time Power Alarm High Differential Pressure Alarm EFMLow Temperature Alarm Low Pressure AlarmLow Supply Volts Alarms Lost Temperature AlarmMemory History Logging Open Door Alarm optionalSoftware Error Alarm Configuration Mode Editing ParametersOptional Keypad and Display Wake-Up on Pulse Event Driven Assigning Function KeysAudit Trail Analog SamplingSpecial Key Combinations Calibration Mode Calibrating the Pressure TransducerCalibrating the Temperature Transducer Zero NEW?Calibrating the Differential Pressure Transmitter EFM Only Optional Equipment Analog Output AO OptionAnalog Output Option MTEK6000 Analog Output Specifications Zero x.xxx 04.000 +y.yyy mA Calibration Enclosure MaintenanceChanging the Battery Mtek Manager PcGas Meter Utility PackagePcGas Customer Monitor DC2000 January Alarms Alarm Code Label I.D Description AddressTable A-5 Standard alarms for MTEK6000 EFC w/ Aux Pressure Label Description Address Aux Pressure Table A-11 Standard alarms for MTEK6000 EFC2 w/2 Aux Press Table A-14 Standard alarms for MTEK6000 EFM w/ Aux Press MTEK6000 EPR MTEK6000 ETR MTEK6000 Eptr January AGA-7 Volume Calculations NX19 Supercompressibility ReportJanuary Audit Trail Logging Enable Alarm Pulse Output Enable/DisableAlarm Pulse Time ms Analog Sampling Rate secondsCalibration Password Base TemperatureCalibrate Mode Time-out Configuration PasswordDate Month, Day, Year Flow Rate Update Interval sCubic Unit/Pulse In or Meter Drive High Pressure Alarm Reset High Flow Rate Alarm SetpointHigh Pressure Alarm Setpoint High Temperature Alarm ResetLow Supply Voltage Alarm Setpoint Low Pressure Alarm SetpointLow Supply Voltage Alarm Reset Low Temperature Alarm ResetWake Up Interval seconds Uncorrected VolumeTime Hhmmss Hours, Minutes, Seconds Wake Up On Pulse event driven EFCUnits of Measure Flow & VolumeTemperature Differential AccumulationsJanuary SBC Revision a Corrector Board OPT Revision a Option Board Description Jumper positionPulse Input #1 PLI Revision C Phone Line Interface Service Requirement Consumer Information and FCC RequirementsWarranty & Repair Service EL0001 Sheet 1 of 2 Class I Division 2 Installation EL0001 Sheet 2 of 2 Class I Division 2 Installation Warranty Information January Appendix G TII Station Protector January Appendix H Hazardous Area Installation Control Drawings October October