MTEK6000 SERIES USER'S MANUAL
61-OPT Revision A – Option Board
|
|
|
Pulse Output 1: | Use | Terminal 19 (normally open contact) |
|
| Terminal 18 (common 1) |
| and | Terminal 17 (normally closed contact) |
| For form A output: put JP1 and JP1A both in position A to B. | |
| For form C output: put JP1 and JP1A both in position B to C (requires external wetting). | |
Pulse Output 2: | Use | Terminal 16 (normally open contact) |
|
| Terminal 15 (common 2) |
| and | Terminal 14 (normally closed contact) |
| For form A output: put JP2 and JP2A both in position A to B. | |
| For form C output: put JP2 and JP2A both in position B to C (requires external wetting). | |
Pulse Output 3 : | Use | Terminal 13 (normally open contact) |
(optional) |
| Terminal 12 (common 3) |
| and | Terminal 11 (normally closed contact) |
| For form A output: put JP3 and JP3A both in position A to B. | |
| For form C output: put JP3 and JP3A both in position B to C (requires external wetting). | |
Pulse Output 4: | Use | Terminal 10 (normally open contact) |
(optional) |
| Terminal 9 (common 4) |
| and | Terminal 8 (normally closed contact) |
| For form A output: put JP4 and JP4A both in position A to B | |
| For form C output: put JP4 and JP4A both in position B to C (requires external wetting). |
JP5 can be used as a convenient means to connect pulse input 2 as follows:
Description | Jumper position |
connect pulse input 2 to mechanical | JP5 A, B, & C all in |
index secondary output |
|
otherwise | JP 5 A, B, & C all out (default) |
January 2002 |