Compaq MTEK6000 Alarm Pulse Output Enable/Disable, Alarm Pulse Time ms, Atmospheric Pressure

Page 59

MTEK6000 SERIES USER'S MANUAL

APPENDIX C:

PARAMETER DESCRIPTION

The parameters relative to the operation and configuration of the MTEK6000 are listed below (See Appendix A for the addresses of these parameters).

Alarm Pulse Output Enable/Disable

The EFC can generate a generic pulse output on any alarm condition. This parameter enables the pulse output through Relay #2. Enter 35.7 to enable or 0.0 to disable. The Default value is 0.

Alarm Pulse Time (ms)

This parameter sets the output band width in milliseconds (ms) for pulse outputs generated by an alarm condition. The user should take care in assigning a value for this parameter. The value should always be less than the Wakeup Interval Parameter; else the unit will stay awake for the duration of the pulse. Range 0 to 32,000 ms. The Default value is 70.

Analog Sampling Rate (seconds)

This parameter sets the rate at which the unit will sample its dynamic analog input variable channels. When enabled, the unit samples pressure, flow temperature, case temperature and the external analog transmitters values once for each selected interval.

The sample’s minimum and maximum values are then checked against their corresponding High and Low Setpoints. If the unit determines that these conditions were exceeded, it “wakes up” and finds the average of all samples accumulated since the last process execution, displays this analog mean, and applies it to any necessary subsequent calculation.

January 2002

To enable analog sampling, set this parameter to 1-99 seconds. Set to 0 to disable analog sampling.

Atmospheric Pressure

If the station is configured for ABSOLUTE

pressure, enter 0. Enter the actual atmospheric pressure for GAUGE pressure. The default value is 14.4 PSI for gauge and 0 for absolute.

Audit Trail Logging Enable

This parameter enables Audit Trail logging in the MTEK6000. Examples of audit trail events include editing any item at the device, and logging the time it enters calibration mode. After the maximum number of events

(250)has been logged, no more parameters can be changed in the unit until the audit trail is uploaded and reset by the software. This parameter is disabled at the factory. The user may enable Audit Trail logging by entering 250. Default value is 0 for disable.

NOTE

Once the audit trail is enabled, the user cannot disable it without reloading the database.

Base Pressure

The Base Pressure parameter appears as a factor in the Corrected Volume calculation. It is one of the factors used to correct the flowing volume, as registered by the meter itself, to the base volume used for calculating “Standard Volume”. The normal pressure ranges for this parameter are 11.000 to 16.000 PSI.

The standard value for this parameter is 14.730 PSIA, 101.560 kPa, or 1.114 kg/cm2, depending upon the system of units chosen for the particular device. Default value is 14.73.

C - 1

Image 59
Contents MTEK6000 Series Table of Contents Optional Equipment List of Figures January January Hazardous Locations OverviewSecurity Options ONE-YEAR WarrantyCompliance MTEK6000 exterior view MTEK6000 interior view Initial Check OUT UnpackingMounting the Instrument on the Meter Power for the MTEK6000Optional Power Supplies Output Shaft Rotation Setting UP the Index AssemblyCounter Masking Unit and Index RotationWall and Pipe Mounting Pulse Input to the MTEK6000Wall Mounting Transducers in the MTEK6000 Connecting the Pressure TubingTypical Installation for MTEK6000 EFC Pipe Size Insertion Thermowell Length Part Number Installing the Thermal ProbePulse Output Wiring Installing the Pulse Output WiringUncorrected Pulse Output Specifications Volume and Alarm Pulse SpecificationsModem Communications 2400 Baud Connecting the Telephone LineCommunications RS-232C Serial Direct Communications 9600 BaudActivity Indicator MTEK6000 Function RS232 cable connected January References GroundingJanuary Display Mode SLEEP/WAKE-UP ModeViewing and clearing alarms using the magnet Alarm ModeHigh Differential Pressure Alarm EFM Faulty Counter Alarm EFCFirst Time Power Alarm High Flow Rate Alarm EFC & EFMLost Temperature Alarm Low Pressure AlarmLow Supply Volts Alarms Low Temperature AlarmMemory History Logging Open Door Alarm optionalSoftware Error Alarm Editing Parameters Configuration ModeOptional Keypad and Display Analog Sampling Assigning Function KeysAudit Trail Wake-Up on Pulse Event DrivenSpecial Key Combinations Calibrating the Pressure Transducer Calibration ModeZero NEW? Calibrating the Temperature TransducerCalibrating the Differential Pressure Transmitter EFM Only Analog Output AO Option Optional EquipmentAnalog Output Option MTEK6000 Analog Output Specifications Zero x.xxx 04.000 +y.yyy mA Calibration Enclosure MaintenanceChanging the Battery Mtek Manager PcGas Meter Utility PackagePcGas Customer Monitor DC2000 January Label I.D Description Address Alarms Alarm CodeTable A-5 Standard alarms for MTEK6000 EFC w/ Aux Pressure Label Description Address Aux Pressure Table A-11 Standard alarms for MTEK6000 EFC2 w/2 Aux Press Table A-14 Standard alarms for MTEK6000 EFM w/ Aux Press MTEK6000 EPR MTEK6000 ETR MTEK6000 Eptr January NX19 Supercompressibility Report AGA-7 Volume CalculationsJanuary Analog Sampling Rate seconds Alarm Pulse Output Enable/DisableAlarm Pulse Time ms Audit Trail Logging EnableConfiguration Password Base TemperatureCalibrate Mode Time-out Calibration PasswordDate Month, Day, Year Flow Rate Update Interval sCubic Unit/Pulse In or Meter Drive High Temperature Alarm Reset High Flow Rate Alarm SetpointHigh Pressure Alarm Setpoint High Pressure Alarm ResetLow Temperature Alarm Reset Low Pressure Alarm SetpointLow Supply Voltage Alarm Reset Low Supply Voltage Alarm SetpointWake Up On Pulse event driven EFC Uncorrected VolumeTime Hhmmss Hours, Minutes, Seconds Wake Up Interval secondsAccumulations Flow & VolumeTemperature Differential Units of MeasureJanuary SBC Revision a Corrector Board Description Jumper position OPT Revision a Option BoardPulse Input #1 PLI Revision C Phone Line Interface Consumer Information and FCC Requirements Service RequirementWarranty & Repair Service EL0001 Sheet 1 of 2 Class I Division 2 Installation EL0001 Sheet 2 of 2 Class I Division 2 Installation Warranty Information January Appendix G TII Station Protector January Appendix H Hazardous Area Installation Control Drawings October October