Optiquest iRIS 220, iRIS 320 Sensor Connection Examples, Connecting a Flow Meter or Rain Gauge

Page 55

53 - iRIS 220 / iRIS 320, V1.19 User Guide

6 Sensor Connection Examples

6.1 Connecting a Flow Meter or Rain Gauge

One of the most common uses for the iRIS is logging data from pulse sources such as flow meters or rain gauges.

Connecting such devices to the iRIS is very simple – wire the switch between the appropriate digital input (DI1 or DI2) and the digital ground (DGND) terminal. Both of the digital inputs provide a “wetting current” for clean contact sources, but transistor switches and active signals (ones that supply a voltage) can also be used. If a transistor switch is used, connect the collector (+) to the digital input and the emitter (-) to the DGND. See section 3.4.5 for details on the digital inputs and setting up the input debounce mode.

Both inputs can be used simultaneously and each input has three associated totalisers, which are viewable from the LCD. See section 5.2.7. These totalisers operate even if the input is not configured as a source to one of the six virtual sensors.

The diagram below shows the typical connection diagram for such an installation. It assumes the use of DI1 as the pulse input channel. The charging source can be any d.c supply from 15V – 30V, including a directly connected solar panel.

TOP

AI4

 

AI3

 

AI2

 

AI1

 

AGND

 

5VOP

 

DI2

 

DI1

 

DO2

 

DO1

 

SDI-12

 

DGND

 

12V+

 

GND (-)

 

VIN+

Charging source e.g solar panel

GND (-)

 

Flow Meter

or

Rain Gauge

BOTTOM

IRIS I/O Connector

The sensor should be configured for the correct channel, scaling and logging regime as described in section

4.3.8.Event mode (sensor mode=2) can be used to reduce the quantity of data logged, especially for rainfall where the actual data density is low.

Image 55
Contents IRIS 220 iRIS Revision History Contents Diagnostic LEDs 13.4 13.113.2 13.3Support IntroductionAbout this Manual Hardware PCB RevisionOverview General Characteristics iRISIntroduction IRIS 320 / iRIS Feature Summary FeaturesTypical Applications Technical Specifications iRIS Technical Specifications iRIS 320 / iRIS Alternative Wireless Connectivity CSD/SMS Key FeaturesTerminal Configuration Wireless IP Connectivity GPRS/CDMAFull Power Save Partial Power SaveData Logging Security Alarm ProcessingLogged Data Array Identification Real Time Clock & CalendarRemoving / fitting the SIM card GSM models only InstallationRemoving/Replacing the Electronic Assembly iRIS Opening / Closing the Housing iRIS 320 / iRISInternal Battery iRIS 320 / iRIS 320V only I/O ConnectorExternal Charger Power Supply Internal / External 12V Battery SupplyAnalog Inputs Digital Input Circuit Digital InputsDigital Output 1 Circuit Digital OutputsMounting iRIS Antenna ConnectionMounting iRIS 320 / iRIS IRIS 320 / iRIS 320V Mounting DiagramTerminal Connection ConfigurationComputer IRIS Terminal Cfg IRIS 220 / iRIS 320, V1.19 User Guide Main Menu Level Terminal MenusSelect this option to display the Comms configuration menu Comms Cfg Level Option 9 SDI-12 Terminal iRIS 320 / iRIS 320V only Press Esc then Enter to enter an empty string GPRS/CDMA-1X Cfg LevelGprs IP Acceptance Cfg Level Phone List LevelGPRS/CDMA-1X Schedule Cfg Level Sensor Selection Level Internal Sensor Sources Log Multiplier 100 100 Sensor Cfg LevelFlag Value Description Mode Name DescriptionSupplementary Logging Flag Definitions Multiplier= Alarm Cfg Level Alarm Selection LevelOutput Cfg Level Output Selection LevelSource 0 1 2 3 4 Description Typical Scheduled Output Example Date/Time Cfg LevelMiscellaneous Menu Level Voice Menu Level Status LED LED IndicatorsDiagnostic LEDs OperationKeypad Buttons LCD & Keypad iRIS 320 onlyLCD Operation Status IconsLevel Level6 Display Menu StructureSystem Status 1 Screen Level Log In Screen Level 0, when not loggedLogging Control Screen Level 0, when logged Primary LCD Display ScreensMain Menu Screen Level Sensor Related ScreensSystem Status 3 Screen Level System Status 4 Screen LevelSensor Settings Screen 4/5 Alarm 1 Level Sensor Settings Screen 1/5 Process LevelSensor Settings Screen 2/5 Scaling Level Sensor Settings Screen 3/5 Logging LevelTotaliser Related Screens Comms Related Screens Comms Setting Screen 5/8 Schedule 1 Level Comms Setting Screen 2/8 APN/Local IP LevelComms Setting Screen 3/8 Socket a Level Comms Setting Screen 4/8 Socket B LevelRS232 Port Mode Screen Level Comms Setting Screen 6/8 Schedule 2 LevelComms Setting Screen 7/8- CSD Settings Level Comms Setting Screen 8/8 RS232 Settings LevelSMS Communication Comms Enable Screen LevelComms Test Screen Level General Hints Iris I/O Connector Connecting a Flow Meter or Rain GaugeSensor Connection Examples Required Connecting a 0-5V Pressure Transducer12V Battery ExternalRequired IRIS I/O Connector Connecting a 2-Wire Loop-Powered 4-20mA SensorSignal 4-20mA Sensor Isolated 12V 24V DC- DC Converter External 12V Battery Optional Connecting an Up/Down Water Level InstrumentConnecting Analogue Wind Instruments Charging source Solar panel Connecting SDI-12 Instruments. iRIS 320 / iRIS 320V onlyTOP AI4 AI3 AI2 AI1 Agnd 5VOP Example a 4-20mA Water Level Sensor Analogue Input ScalingTroubleshooting User Notes DS-IRM Modem DB25 M Appendix a Radio Using the RS232 InterfaceTait TM80xx Radio DB15 MiRIS 3xx / DS-4483 DB9 F IRIS 320 DB9 FPartition Appendix B Voice Annunciation iRISAppendix C SDI-12 iRIS 320 / 320V only What is SDI-12?Advantages of SDI-12 Volt-Line SDI-12 Electrical InterfaceSerial Data Line Ground LineAppendix D Legacy PCB Rev 1.1 Details 13.1 I/O ConnectorDebounce Links Interface Wire Terminal SDI-12 Bus Function Diagnostic LEDsSDI-12 Interface REDSupply Voltage Transistor SwitchPurple Switched Output Switch ModuleApproximate Mechanical Properties Appendix E iRIS 320 Enclosure Material SpecChemical Composition IQuest Global Data Network IQuest APN