Optiquest iRIS 320, iRIS 220 manual Multiplier=

Page 36

iRIS 220 / iRIS 320, V1.19 User Guide - 34

Option 4

When this option is selected you will be prompted to enter a scaling multiplier. This multiplier is used to convert the raw input into engineering units.

> Multiplier=

Option 5

When this option is selected you will be prompted to enter a scaling offset. This offset is added to the scaled engineering value.

> Offset=

Option 6

When this option is selected you will be prompted to enter a logging multiplier to convert from engineering units to an integer value to storage in the logging memory.

> Log Multiplier=

For example, if you need to log a measurement that is accurate to two decimal places you will need to enter a logging multiplier of 100.

NOTE: Care needs to be taken in the selection of an appropriate logging multiplier because the iRIS logs data as signed 16-bit integer values (-32768 to 32767). This means that the maximum scaled value multiplied by the logging multiplier must not exceed 32767.

Option 7

When this option is selected you will be prompted to enter a logging rate (in minutes) for the sensor.

> Log Rate=

If you wish to log digital data on change of state you can enter a value of 0. If this parameter is left at 0 for Analog sources, they will not be logged.

Option 8

Select this option to display the Alarm Configuration menu for the sensor.

Option 9

Select this information to view the logged data for the sensor. You will be prompted to enter the number of samples you would like returned

NOTE: This option is only available when the terminal session is through the RS232 port, not through a GPRS/CDMA-1X terminal session.

34

iQuest (NZ) Ltd - PO Box 15169, Hamilton, New Zealand Tel: +64 7 857-0810 Fax: +64 7 857-0811 Email: iquest@iquest.co.nz

Image 36
Contents IRIS 220 iRIS Revision History Contents Diagnostic LEDs 13.1 13.213.3 13.4Introduction About this ManualHardware PCB Revision SupportGeneral Characteristics iRIS OverviewIntroduction IRIS 320 / iRIS Features Feature SummaryTypical Applications Technical Specifications iRIS Technical Specifications iRIS 320 / iRIS Key Features Terminal ConfigurationWireless IP Connectivity GPRS/CDMA Alternative Wireless Connectivity CSD/SMSPartial Power Save Full Power SaveData Logging Alarm Processing Logged Data Array IdentificationReal Time Clock & Calendar SecurityInstallation Removing/Replacing the Electronic Assembly iRISOpening / Closing the Housing iRIS 320 / iRIS Removing / fitting the SIM card GSM models onlyI/O Connector Internal Battery iRIS 320 / iRIS 320V onlyInternal / External 12V Battery Supply External Charger Power SupplyAnalog Inputs Digital Inputs Digital Input CircuitDigital Outputs Digital Output 1 CircuitAntenna Connection Mounting iRISIRIS 320 / iRIS 320V Mounting Diagram Mounting iRIS 320 / iRISConfiguration Terminal ConnectionComputer IRIS Terminal Cfg IRIS 220 / iRIS 320, V1.19 User Guide Terminal Menus Main Menu LevelSelect this option to display the Comms configuration menu Comms Cfg Level Option 9 SDI-12 Terminal iRIS 320 / iRIS 320V only GPRS/CDMA-1X Cfg Level Press Esc then Enter to enter an empty stringGprs Phone List Level IP Acceptance Cfg LevelGPRS/CDMA-1X Schedule Cfg Level Sensor Selection Level Sensor Cfg Level Internal Sensor Sources Log Multiplier 100 100Mode Name Description Flag Value DescriptionSupplementary Logging Flag Definitions Multiplier= Alarm Selection Level Alarm Cfg LevelOutput Selection Level Output Cfg LevelSource 0 1 2 3 4 Description Date/Time Cfg Level Typical Scheduled Output ExampleMiscellaneous Menu Level Voice Menu Level LED Indicators Diagnostic LEDsOperation Status LEDLCD & Keypad iRIS 320 only LCD OperationStatus Icons Keypad ButtonsDisplay Menu Structure Level Level6Log In Screen Level 0, when not logged Logging Control Screen Level 0, when loggedPrimary LCD Display Screens System Status 1 Screen LevelSensor Related Screens System Status 3 Screen LevelSystem Status 4 Screen Level Main Menu Screen LevelSensor Settings Screen 1/5 Process Level Sensor Settings Screen 2/5 Scaling LevelSensor Settings Screen 3/5 Logging Level Sensor Settings Screen 4/5 Alarm 1 LevelTotaliser Related Screens Comms Related Screens Comms Setting Screen 2/8 APN/Local IP Level Comms Setting Screen 3/8 Socket a LevelComms Setting Screen 4/8 Socket B Level Comms Setting Screen 5/8 Schedule 1 LevelComms Setting Screen 6/8 Schedule 2 Level Comms Setting Screen 7/8- CSD Settings LevelComms Setting Screen 8/8 RS232 Settings Level RS232 Port Mode Screen LevelComms Enable Screen Level SMS CommunicationComms Test Screen Level General Hints Connecting a Flow Meter or Rain Gauge Iris I/O ConnectorSensor Connection Examples Connecting a 0-5V Pressure Transducer 12V BatteryExternal RequiredConnecting a 2-Wire Loop-Powered 4-20mA Sensor Required IRIS I/O ConnectorSignal 4-20mA Sensor Isolated 12V 24V DC- DC Converter Connecting an Up/Down Water Level Instrument External 12V Battery OptionalConnecting Analogue Wind Instruments Connecting SDI-12 Instruments. iRIS 320 / iRIS 320V only Charging source Solar panelTOP AI4 AI3 AI2 AI1 Agnd 5VOP Analogue Input Scaling Example a 4-20mA Water Level SensorTroubleshooting User Notes Appendix a Radio Using the RS232 Interface Tait TM80xx Radio DB15 MiRIS 3xx / DS-4483 DB9 FIRIS 320 DB9 F DS-IRM Modem DB25 MAppendix B Voice Annunciation iRIS PartitionWhat is SDI-12? Appendix C SDI-12 iRIS 320 / 320V onlyAdvantages of SDI-12 SDI-12 Electrical Interface Serial Data LineGround Line Volt-Line13.1 I/O Connector Appendix D Legacy PCB Rev 1.1 DetailsDebounce Links Diagnostic LEDs Interface Wire Terminal SDI-12 Bus FunctionSDI-12 Interface Transistor Switch Purple Switched OutputSwitch Module REDSupply VoltageAppendix E iRIS 320 Enclosure Material Spec Approximate Mechanical PropertiesChemical Composition IQuest APN IQuest Global Data Network