2.1.5 Electrical Installation
2.1.5.1 Power Connections
Warning: Do not connect ac power to your meter until you have completed all output connections. This meter must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!
Connect the main power connections as shown below.
90 to 240 Vac 110 to 375 Vdc
OR
Use copper conductors only for power connections
24 Vac or
20 to 36 Vdc
Figure 2.7 Main Power Connections Table 2.3 Fuse Requirement (See Specifications Section)
FUSE | Connector | Output Type | For 115Vac | For 230Vac | DC |
FUSE 1 | Power | N/A | 100 mA(T) | 100 mA(T) | 100 mA(T) |
FUSE 2 | Power | N/A | N/A | N/A | 400 mA(T) |
For the low voltage power option, in order to maintain the same degree of protection as the standard high voltage input power units (90 - 240 Vac), always use a Safety Agency Approved DC or AC source with the same Overvoltage Category and pollution degree as the standard AC unit (90 - 240 Vac).
The Safety European Standard