2.1.5 Electrical Installation

2.1.5.1 Power Connections

Warning: Do not connect ac power to your meter until you have completed all output connections. This meter must only be installed by a specially trained electrician with corresponding qualifications. Failure to follow all instructions and warnings may result in injury!

Connect the main power connections as shown below.

90 to 240 Vac 110 to 375 Vdc

OR

Use copper conductors only for power connections

24 Vac or

20 to 36 Vdc

Figure 2.7 Main Power Connections Table 2.3 Fuse Requirement (See Specifications Section)

FUSE

Connector

Output Type

For 115Vac

For 230Vac

DC

FUSE 1

Power

N/A

100 mA(T)

100 mA(T)

100 mA(T)

FUSE 2

Power

N/A

N/A

N/A

400 mA(T)

For the low voltage power option, in order to maintain the same degree of protection as the standard high voltage input power units (90 - 240 Vac), always use a Safety Agency Approved DC or AC source with the same Overvoltage Category and pollution degree as the standard AC unit (90 - 240 Vac).

The Safety European Standard EN61010-1 for measurement, control, and laboratory equipment requires that fuses must be specified based on IEC127. This standard specifies for a Time-lag fuse, the letter code “T”. The above recommended fuses are of the type IEC127-2-sheet III. Be aware that there are significant differences between the requirements listed in the UL 248-14/CSA 248.14 and the IEC 127 fuse standards. As a result, no single fuse can carry all approval listings. A 1.0 Amp IEC fuse is approximately equivalent to a 1.4 Amp UL/CSA fuse. It is advised to consult the manufacturer’s data sheets for a cross-reference.

11

Image 17
Contents User’sGuide Page Table of Contents List of Figures 19a List of Tables Part Introduction Safety ConsiderationsBefore You Begin Customer ServiceDescription Inspecting Your ShipmentDescription Rockwell Automation, and National Instruments, among others Front Panel Enunciators Physical Characteristics and Mounting 2.1.1 Front PanelNetwork Rear Panel ConnectorRear Panel Mounting Dimensions DimensionsPanel Mounting Assembly and Mounting 2.1.4.1 Panel Mounting InstructionAntenna Mounting Antenna Mounting InstructionAccessing the Main Board Assembly Disassembly InstructionUse copper conductors only for power connections Electrical Installation Power ConnectionsOutputs Wiring Hookup Wiring OutputsOutputs Wiring Hookup Pulse and AnalogFast Blow Typical Applications10 Location of S1 Jumper Jumper Settings for Display Color SetupPosition Dip SwitchesPin Name Description Network Communication Interfaces 1 10Base-T RJ-45 PinoutDhcp Ethernet MAC AddressNetwork Protocols .1 Label Detail DNS 4 Position DIP SwitchIP Address Default IP AddressSection Changing TCP/IP Properties on Your ComputerPinging the Meter from MS-DOS Prompt Testing the ConnectionIConnect Software Assigning an IP Address using iConnectWireless Sensor Home Access Control Setting a New IP Address over the NetworkWireless System Home Page Menu Meter’s Configurations and OperationsPower Up Meter Login AdministratorLab Get Readings from the End DeviceGet Readings from the End Device ID End Device ID/address Java 1.5.x.x. Screen Shots Java Runtime Environment Setup10 Java 1.4.2.x Screen Shot Java Runtime Environment 1.4 Setup InstructionsAccessing Meter units using the internet Accessing Meter units within your internal networkDiagnostics Accessing Meter units over Peer-to-Peer network11 Java Policy Java Policy12 Java Policy Djava.security.policy=C\0JAVAPOLICY\javapolicy.txt13 Chart ChartChart 14 Controller Setup Controller SetupTransmission Power SetupMonitor MiscellaneousInput SetupSetpoints & On/Off Control Setpoints & On/Off Control Alarms 1 Setup Analog Output Retransmission Setup19a Display Menu Display19b Display Menu if Analog Output Option Example Display Color ExamplesAmber 20 ID Code Passcode IDRemote Access Tunneling Network SetupGeneral Terminal ServerTerminal Server Remote AccessGeneral 22 End Device Setup End Device SetupEnd Device Parameters Sensor2 Sensor1T75F T74F SensorDisplay UpdateCancel Access Control Command Description Response Telnet SetupList of Ascii / Telnet Commands 01 indicates End Device #1 HTTPget Program HTTPget using PortHTTPget and ARP to Setup Device IP Address Connector/Transmitter TypesARP Protocol 27 ARP Commands and Responses28 iLog Software Logging Data for End Device #2, 6 ILog SoftwareILog Software ILog Excel Applications Wireless DevicesNetworked Product ILog File Alternate Networked Product Auto FullMail Notifier Software InstallationEmail Setup Email Address SetupMail Notifier Commands Device Setting and ConfigurationAlarm Editor General Deployment Guidelines Environment / Operating ConditionsOperation in Buildings Distance feet With Line-of-SightCasing and Closure Around the Antenna Without Line-of-SightMaterial Type See .1.4.2 and FigurePart Specifications General Safety & Regulatory Compliance Input Factory Preset ValuesPasscode ID Factory Preset Values Setpoints & On/Off ControlAlarm Remote Access Tunneling Factory Preset Values Analog Output RetransmissionAccess Control MiscEnd Device # 1 Factory Preset Values End Device SetupAppendix Aglossary Appendix BIP Address Address Class Mask Binary Value Mask Decimal Value Ascii Dec Hex Ctrl Key Ascii Control Codes Definition Appendix E ILog Error Messages Error # Description Maximum Power Levels and Antenna Gains WARRANTY/DISCLAIMER PH/CONDUCTIVITY Shop on line at omega.com