Omega WI8XX-U manual HTTPget Program, HTTPget using Port

Page 64

4.5 HTTPget Program

The Httpget software is used to send a single HTTP or TCP request to the Meter. In contrast, the telnet or Hyperterminal programs allow a continuous connection with multiple requests to be sent to the Meter.

Generally HTTPget is used for simply programming an IP address to the Meter or for quickly obtaining a reading from a End Device.

The Meter must be configured from the "Network Setup" page so that the "TCP Connections" indicates any number between 1 and 5. Make sure that the Port Field number indicates 2000.

Whenever Terminal Server service (using Port 2000 by default) is required, the # of connections must be set to a value from 1 to 5. The Terminal Server mode is the recommended mode for the most reliable connection when operating with NEWPORT software or with other programs supporting TCP/IP communications. The Port 2000 access can be used with NEWPORT software and may be needed with some Meters when you need to view readings from the web page while simultaneously collecting data through TCP/IP communications.

4.5.1 HTTPget using Port 2000

You can setup and read the information from the Meter by using the HTTPget program. The following program can be used to read data from the embedded server firmware by using TCP port 2000. The command string is sent to this TCP port, then the response can be read back from the same socket.

The HTTPget.exe file is used to setup and read information from the Meter. This file will be automatically installed when you install the MailNotifier software available on our website and CD.

Notes on using HTTPget :

The Httpget.exe program is installed to the windows directory (usually c:\winnt or c:\windows) when installing the Mail Notifier software.

1). Open up a command window (or have a DOS window) a) Click on start menu

b) Click on "Run"

c) In the dialog box that appears, type "cmd" or "command" and click on "OK" button.

d) A command window should now appear.

2). If you now type "httpget" and press the "enter", the program options should be displayed.

3). Next run HTTPget with the options displayed below httpget -r-S *SR00z 192.168.1.135:2000 -C1-q

58

Image 64
Contents User’sGuide Page Table of Contents List of Figures 19a List of Tables Safety Considerations Part IntroductionCustomer Service Before You BeginDescription Inspecting Your ShipmentDescription Rockwell Automation, and National Instruments, among others Physical Characteristics and Mounting 2.1.1 Front Panel Front Panel EnunciatorsRear Panel Rear Panel ConnectorNetwork Dimensions Mounting DimensionsAssembly and Mounting 2.1.4.1 Panel Mounting Instruction Panel MountingAntenna Mounting Instruction Antenna MountingDisassembly Instruction Accessing the Main Board AssemblyElectrical Installation Power Connections Use copper conductors only for power connectionsWiring Outputs Outputs Wiring HookupOutputs Wiring Hookup Pulse and AnalogTypical Applications Fast BlowJumper Settings for Display Color Setup 10 Location of S1 JumperDip Switches PositionNetwork Communication Interfaces 1 10Base-T RJ-45 Pinout Pin Name DescriptionNetwork Protocols .1 Label Detail Ethernet MAC AddressDhcp 4 Position DIP Switch DNSDefault IP Address IP AddressChanging TCP/IP Properties on Your Computer SectionTesting the Connection Pinging the Meter from MS-DOS PromptAssigning an IP Address using iConnect IConnect SoftwareWireless Sensor Home Setting a New IP Address over the Network Access ControlMeter’s Configurations and Operations Wireless System Home Page MenuLogin Administrator Power Up MeterGet Readings from the End Device LabGet Readings from the End Device ID End Device ID/address Java Runtime Environment Setup Java 1.5.x.x. Screen ShotsJava Runtime Environment 1.4 Setup Instructions 10 Java 1.4.2.x Screen ShotAccessing Meter units within your internal network Accessing Meter units using the internetDiagnostics Accessing Meter units over Peer-to-Peer networkJava Policy 11 Java PolicyDjava.security.policy=C\0JAVAPOLICY\javapolicy.txt 12 Java PolicyChart 13 ChartChart Controller Setup 14 Controller SetupSetup Transmission PowerMonitor MiscellaneousSetup InputSetpoints & On/Off Control Setpoints & On/Off Control Alarms 1 Setup Setup Analog Output RetransmissionDisplay 19a Display Menu19b Display Menu if Analog Output Option Display Color Examples ExampleAmber Passcode ID 20 ID CodeNetwork Setup Remote Access TunnelingGeneral Terminal ServerGeneral Remote AccessTerminal Server End Device Setup 22 End Device SetupEnd Device Parameters Sensor1 Sensor2Sensor T75F T74FCancel UpdateDisplay Access Control List of Ascii / Telnet Commands 01 indicates End Device #1 Telnet SetupCommand Description Response HTTPget using Port HTTPget ProgramConnector/Transmitter Types HTTPget and ARP to Setup Device IP Address27 ARP Commands and Responses ARP ProtocolILog Software 28 iLog Software Logging Data for End Device #2, 6Wireless Devices ILog Software ILog Excel ApplicationsNetworked Product ILog File Alternate Networked Product Auto FullInstallation Mail Notifier SoftwareEmail Address Setup Email SetupDevice Setting and Configuration Mail Notifier CommandsAlarm Editor Environment / Operating Conditions General Deployment GuidelinesOperation in Buildings With Line-of-Sight Distance feetWithout Line-of-Sight Casing and Closure Around the AntennaMaterial Type See .1.4.2 and FigurePart Specifications General Safety & Regulatory Compliance Factory Preset Values InputAlarm Factory Preset Values Setpoints & On/Off ControlPasscode ID Factory Preset Values Analog Output Retransmission Remote Access TunnelingAccess Control MiscFactory Preset Values End Device Setup End Device # 1Appendix Aglossary Appendix BIP Address Address Class Mask Binary Value Mask Decimal Value Ascii Dec Hex Ctrl Key Ascii Control Codes Definition Appendix E ILog Error Messages Error # Description Maximum Power Levels and Antenna Gains WARRANTY/DISCLAIMER Shop on line at omega.com PH/CONDUCTIVITY