Omega WI8XX-U manual Factory Preset Values, Input

Page 80

PART 7

FACTORY PRESET VALUES

To set the Wireless and Ethernet settings back to Factory Default do the following to the DIP switches (see Figure 2.11).

To set the Ethernet board to Factory Default settings:

1) Put the 4 position DIP switch #2 to ON position (it does not matter if the Meter is On or Off).

2) Power-cycle the Meter and wait about 10 seconds until the Meter fully boots up.

3) Put the DIP switch #2 back to OFF position (it does not matter if the Meter is powered On or Off, just make sure that the DIP switch is set to OFF, otherwise, everytime the unit is power-cycled the factory settings will take over.

To set the Wireless board to Factory Default settings:

1) Put the 8 position DIP switch #1 to ON position (it does not matter if the Meter is On or Off).

2) Power-cycle the Meter and wait about 10 seconds until the Meter fully boots up.

3) Put the DIP switch #1 back to OFF position (it does not matter if the Meter is powered On or Off, just make sure that the DIP switch is set to OFF, otherwise, everytime the unit is power-cycled the factory settings will take over.

The defaults that will change are noted by a "W" or "E" in the Table below. The defaults for the Meter/Controller "C" are set by the individual Setup menus.

C = Meter/Controller Board

W = Wireless or Coordinator Board

E = Ethernet or iServer Board

Table 7.1 Factory Preset Values

CONTROLLER SETUP

 

Input

1

Control End Device ID

Filter

4

Temperature Unit

F

W

C C

74

Image 80
Contents User’sGuide Page Table of Contents List of Figures 19a List of Tables Safety Considerations Part IntroductionCustomer Service Before You BeginDescription Inspecting Your ShipmentDescription Rockwell Automation, and National Instruments, among others Physical Characteristics and Mounting 2.1.1 Front Panel Front Panel EnunciatorsNetwork Rear Panel ConnectorRear Panel Dimensions Mounting DimensionsAssembly and Mounting 2.1.4.1 Panel Mounting Instruction Panel MountingAntenna Mounting Instruction Antenna MountingDisassembly Instruction Accessing the Main Board AssemblyElectrical Installation Power Connections Use copper conductors only for power connectionsWiring Outputs Outputs Wiring HookupOutputs Wiring Hookup Pulse and AnalogTypical Applications Fast BlowJumper Settings for Display Color Setup 10 Location of S1 JumperDip Switches PositionNetwork Communication Interfaces 1 10Base-T RJ-45 Pinout Pin Name DescriptionDhcp Ethernet MAC AddressNetwork Protocols .1 Label Detail 4 Position DIP Switch DNSDefault IP Address IP AddressChanging TCP/IP Properties on Your Computer SectionTesting the Connection Pinging the Meter from MS-DOS PromptAssigning an IP Address using iConnect IConnect SoftwareWireless Sensor Home Setting a New IP Address over the Network Access ControlMeter’s Configurations and Operations Wireless System Home Page MenuLogin Administrator Power Up MeterGet Readings from the End Device LabGet Readings from the End Device ID End Device ID/address Java Runtime Environment Setup Java 1.5.x.x. Screen ShotsJava Runtime Environment 1.4 Setup Instructions 10 Java 1.4.2.x Screen ShotAccessing Meter units within your internal network Accessing Meter units using the internetDiagnostics Accessing Meter units over Peer-to-Peer networkJava Policy 11 Java PolicyDjava.security.policy=C\0JAVAPOLICY\javapolicy.txt 12 Java PolicyChart 13 ChartChart Controller Setup 14 Controller SetupSetup Transmission PowerMonitor MiscellaneousSetup InputSetpoints & On/Off Control Setpoints & On/Off Control Alarms 1 Setup Setup Analog Output RetransmissionDisplay 19a Display Menu19b Display Menu if Analog Output Option Display Color Examples ExampleAmber Passcode ID 20 ID CodeNetwork Setup Remote Access TunnelingGeneral Terminal ServerTerminal Server Remote AccessGeneral End Device Setup 22 End Device SetupEnd Device Parameters Sensor1 Sensor2Sensor T75F T74FDisplay UpdateCancel Access Control Command Description Response Telnet SetupList of Ascii / Telnet Commands 01 indicates End Device #1 HTTPget using Port HTTPget ProgramConnector/Transmitter Types HTTPget and ARP to Setup Device IP Address27 ARP Commands and Responses ARP ProtocolILog Software 28 iLog Software Logging Data for End Device #2, 6Wireless Devices ILog Software ILog Excel ApplicationsNetworked Product ILog File Alternate Networked Product Auto FullInstallation Mail Notifier SoftwareEmail Address Setup Email SetupDevice Setting and Configuration Mail Notifier CommandsAlarm Editor Environment / Operating Conditions General Deployment GuidelinesOperation in Buildings With Line-of-Sight Distance feetWithout Line-of-Sight Casing and Closure Around the AntennaMaterial Type See .1.4.2 and FigurePart Specifications General Safety & Regulatory Compliance Factory Preset Values InputPasscode ID Factory Preset Values Setpoints & On/Off ControlAlarm Factory Preset Values Analog Output Retransmission Remote Access TunnelingAccess Control MiscFactory Preset Values End Device Setup End Device # 1Appendix Aglossary Appendix BIP Address Address Class Mask Binary Value Mask Decimal Value Ascii Dec Hex Ctrl Key Ascii Control Codes Definition Appendix E ILog Error Messages Error # Description Maximum Power Levels and Antenna Gains WARRANTY/DISCLAIMER Shop on line at omega.com PH/CONDUCTIVITY