4.8.3 Device Setting and Configuration
The Meter and End Devices should first be configured and ready to use.
Make sure to have the following settings in the "Network" web page of the Meter (Figure 4.20) .
Number of Connections = 5
Port = 2000 (other values may be acceptable as long as Mail Notifier is setup with the same Port number)
Then on the Mail Notifier Alarm Editor, set the BusAddress/DeviceID to match the Device ID number for an End Device and set the Reading Cmd to a Command String from the following table:
| Table 4.4 Mail Notifier Commands |
|
|
| ||
| Desired Value to Monitor |
| Reading Command |
| Actual Command |
|
|
|
|
| |||
|
|
| String |
| String (see Note 2) |
|
| Temperature1 (Process) |
| SR##a |
| *SR05a\r |
|
| in scale configured at Meter |
|
|
|
|
|
| Temperature1 (Process) |
| SR##o |
| *SR05o\r |
|
| in Degree C |
|
|
|
|
|
| Temperature2 (Ambient at |
| SR##b |
| *SR05b\r |
|
| End Device sensor) in scale |
|
|
|
|
|
| configured at Meter |
|
|
|
|
|
| Temperature2 (Ambient at |
| SR##p |
| *SR05p\r |
|
| End Device sensor) in Degree C |
|
|
|
|
|
| Battery voltage (mV) |
| SR##j |
| *SR05j\r |
|
| Signal Level |
| SR##h |
| *SR05h\r |
|
| Success Rate |
| SR##i |
| *SR05i\r |
|
1.The commands used in Mail Notifier are symbolic commands rather than actual Meter commands described in Table 4.1.
2.The sample commands in the table above are shown for a sensor with Device ID of 5 such that SR##a, for example, is sent as *SR05a\r.