Soehnle 7731, 7730 user manual Classification, Safety information, Application part

Page 21

4.3 Classification

Medical product of class I with measuring function.

Electrical Safety Class II (safety isolated, no protective conductor connection). Water protection according to EN60529: IPX2 for the whole device.

Application part

The medical product is a Type B application part and intended for direct contact with the patient. The leakage currents correspond to the classification of type B application parts.

4.4 Safety information

This device is designed for professional use. Please read carefully the information listed in the user manual before using the device. It contains important information regarding installation, intended use and maintenance of the device. The manufacturer is not liable

if the following instructions are not followed: When using electrical components under increased safety requirements, the relevant provisions are complied with. Never work on the device, if it is energized.

There are no special requirements for the staff, the scale can be used anywhere. The use in critical environments is not provided (CT, defibrillator). The product is not for use in explosion- prone environments. The scale with mains operation must be set up, so that the separation of power supply and scale can be easily done (ensure accessibility of the socket). The display and the scale may not get wet from below.

Raise the scale in order to move it; do not drag it along the ground. Afterwards check the levelling and adjust if necessary.

In accordance with EC guideline 2004/108/EC, this device is equipped with interference suppression. However, the display reading can be affected by extreme electrostatic or electromagnetic interference, for example if a radio or mobile phone is used immediately beside it. The display should return to normal once the source of interference is removed. Possibly a restart of the scale is necessary. If there is permanent electrostatic interference in the area, please contact your responsible service partner.

This equipment is a measuring instrument. Air draught, vibrations, rapid temperature changes and direct sunlight may affect the weight results. The scale conforms to protection class IPX2. Avoid high levels of air humidity, steam, aggressive fluids and heavy soiling.

21

www.soehnle-professional.com

Image 21
Contents 7730 Page Funktionen WAAGEN-MODELLEAufstellen der Waage und Inbetriebnahme EinschaltenTarieren WiegenLängenmesser mit BMI-Anzeige Kleinkind-VerwiegungAnzeige Ursachen Beseitigung Störungen Ursachen UND BeseitigungAllgemeine Hinweise Technische DatenVerwendungszweck Klassifizierung SicherheitshinweisAnwendungsteil Wartung und Service ReinigungFolgende Desinfektionsmittel sind erlaubt Batterie- und Akku-Entsorgung Gewährleistung / HaftungErläuterung der Symbole Service-SchnittstelleEichung Entsorgung der WaageErläuterung der Symbole auf der Verpackung Erläuterung des Eich-Etiketts und SymboleWarnhinweise Hinweis ZUR Elektromagnetischen Verträglichkeit Störfestigkeitsprüfung IEC 60601-Prüfpegel IEC 60601-Prüfpegel Schutzabstand abhängig von der Sendefrequenz Display with rechargeable battery operation Switching onModels of Scales FunctionsWeighing Switching off / End the operation safelyTaring Height rod with BMI-displayDisplay Causes Remedy Faults Causes and RemediesTechnical data General InformationIntended use Safety information ClassificationApplication part Cleaning Maintenance and serviceFollowing disinfectants are allowed Disposal of batteries Warranty / LiabilityApproval Service interfaceDisposal of scale EC-Mark of conformityExplanation of the approval label and symbols Explanation of the symbols on the packagingExplanation of the symbols Page Modeles DE Balances Installation de la balance et la mise en serviceFonctions Afficheur avec batterieDémarrage Arrêt / Finir le service sûrePesage TarageConsignes Generales Defauts Causes ET SolutionsPage Page Soehnle Industrial Solutions GmbH