![](/images/new-backgrounds/1293430/29343053x1.webp)
Pressure Sensor S/W Error (PE)
Is the Connector connected correctly to the Main PCB and the Pressure Switch ? Or is the Harness alright ?
YES
Is the resistance of the Pressure Switch out of range?
[Pin1 ~ Pin3]
21~23Ω)
NO
Is the air chamber and the tube clogged?
NO
Replace the Main PCB.
NO
YES ![](/images/new-backgrounds/1293430/29343053xi4.webp)
YES
Reconnect or repair the Connector.
Or replace the Harness.
Replace the Pressure Switch.
Check air chamber and remove the foreign material.
26