5 Maintenance

5.1General

User maintenance on this instrument is limited to checking the LED indicator, a diagnostic mode and cleaning the case.

5.2LED Indicator

With the lid of the instrument removed, an LED located on the PCB flashes when a pressure conversion is carried out. This indicates that the instrument has power, that the processor is functioning and the transducer frequency and diode voltage are within normal limits.

5.3Diagnostic Mode

D - Diagnostic Mode

This command places the instrument into the diagnostic mode.

Command: D, mode

Where:

mode is a single digit:

1Dump all 512 stored values.

2Dump transducer coefficients (checks LOAD L command).

3Dump calibration factors i.e. gain, offset etc.

4Continuous output of sensor diode voltage (temperature) mV.

5Continuous output of sensor frequency in Hz.

Result: The reply depends on the mode as follows:

Mode 1

32 lines of output as follows:

aaa- nn nn nn nn nn nn nn nn nn nn nn nn nn nn nn nn<RT> where: aaa = hexadecimal address offset (0 to 1FFH) and

nn = hexadecimal data

K177 Issue No. 3

Image 34
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Description SpecificationDimensions AccuracyData Communications Power supplyPressure measurement Pressure connectionAccessories Mounting the Instrument Figure Pressure ConnectionInstrument Dimensions Electrical Connection Communication connectionsConnections for remote on/off control Communications Changing the communication interface standard FigureLK1 LK2 PL3PL1 RPT ScreeningCommand Summary System OperationSet-up Command Function MeasurementCR Crlf Command Terminators User calibrationSingle commands General Command FormatCommand Strings Page Error Message General Reply FormatData out of range Get reading Detailed Command Definitions Measurement CommandsReturn stored reading Auto send readingAuto Send from switch-on Set-up Commands Unit Set-upSet Resolution Command F, step, averageFilter Set-up Set Communications Options Status Command I,S/PIN, Name ValueCR formatIdentity  Calibration Check User Calibration Linearity/Drift CompensationPIN Change PIN PIN, New PINDiagnostic Mode LED IndicatorDiagnostic Mode GeneralCleaning Approved Service Agents