22 | RPT 301 User Manual |
S - Status
This command returns the instrument status.
Command: | S |
|
Reply: | STATUS nn | |
Where nn: | 01 | = Frequency outside limits |
| 02 | = Diode voltage outside limits |
| 04 | = Electronics failure |
| Combination errors: a status 06 represents diode |
| voltage and electronics failure. |
Result: | no change |
I - Identity |
|
This command requests the unit to send items of data which describe the unit identity.
Command: | I |
|
|
Reply: | type, range, S/N, cal date | ||
Result: | No change in status or operation | ||
Where: | Type | = | Equipment model (Druck RPT 301) |
| Range | = | Pressure range, lower and upper and |
|
|
| pressure type |
| S/N | = | Transducer serial number |
| Cal date = | Date of last calibration |
Command: I,S/PIN, Name Value<CR> format
i.When Name = S, Value =
Result: sets displayed serial number
ii.When Name = H, Value = nnnn Reply: HIGH RANGE = nnn mbar
Result: sets displayed highest operating pressure (mbar)
iii.When Name = L, Value = nnnn Reply: LOWEST RANGE = nnn mbar
Result: sets displayed lowest operating pressure (mbar)