20 | RPT 301 User Manual |
F - Filter Set-up
This command sets the instrument reading filter coefficients.
Command: F, step, average
Where: step is a pressure band within which the filter operates.
If the pressure steps outside this band, the filter is bypassed. This gives an unfiltered response to
step takes a value expressed as a percentage (0 - 100%) of
If step = 0 then the filter is disabled.
average is the number of sequential readings (1 -
99)averaged by the filter. High values effectively give the filter a long
Reply: None
B - Set Resolution
This command enables the decimal place of the reading to be adjusted within the instrument’s resolution, see specification. The reading resolution is automatically selected according to the instrument full- scale and the units that have been selected using the units
Changing units cancels any previous set resolution command.
The command is nonvolatile and remains selected until the units are changed. Changing and restoring the units can be used to cancel the set resolution command.
Command: B,n
Reply: None
Result: Sets the transmitted number of decimal places to be ‘n’