GE RPT 301 user manual Filter Set-up, Command F, step, average, Set Resolution

Page 28

20

RPT 301 User Manual

F - Filter Set-up

This command sets the instrument reading filter coefficients.

Command: F, step, average

Where: step is a pressure band within which the filter operates.

If the pressure steps outside this band, the filter is bypassed. This gives an unfiltered response to non-noise step changes.

step takes a value expressed as a percentage (0 - 100%) of full-scale.

If step = 0 then the filter is disabled.

average is the number of sequential readings (1 -

99)averaged by the filter. High values effectively give the filter a long time-constant.

Reply: None

B - Set Resolution

This command enables the decimal place of the reading to be adjusted within the instrument’s resolution, see specification. The reading resolution is automatically selected according to the instrument full- scale and the units that have been selected using the units set-up command. An algorithm is used that gives a resolution which does not interfere with the accuracy of measurement and is as noise-free as possible before filtering. The set resolution command allows the number of decimal places of the transmitted value to be specified, this overrides the automatic selection of resolution for the current selected units.

Changing units cancels any previous set resolution command.

The command is nonvolatile and remains selected until the units are changed. Changing and restoring the units can be used to cancel the set resolution command.

Command: B,n

Reply: None

Result: Sets the transmitted number of decimal places to be ‘n’ 0-5 max depends on units resolution.

K177 Issue No. 3

Image 28
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Specification DimensionsDescription AccuracyPower supply Pressure measurementData Communications Pressure connectionAccessories Mounting the Instrument Figure Pressure ConnectionInstrument Dimensions Electrical Connection Communication connectionsConnections for remote on/off control Communications Changing the communication interface standard FigureLK1 LK2 PL3PL1 RPT ScreeningCommand Summary System OperationCommand Function Measurement CR Crlf Command TerminatorsSet-up User calibrationSingle commands General Command FormatCommand Strings Page Error Message General Reply FormatData out of range Detailed Command Definitions Measurement Commands Return stored readingGet reading Auto send readingAuto Send from switch-on Set-up Commands Unit Set-upSet Resolution Command F, step, averageFilter Set-up Set Communications Options Status Command I,S/PIN, Name ValueCR formatIdentity  Calibration Check User Calibration Linearity/Drift CompensationPIN Change PIN PIN, New PINLED Indicator Diagnostic ModeDiagnostic Mode GeneralCleaning Approved Service Agents