GE RPT 301 user manual General Reply Format, Error Message

Page 23

RPT 301 User Manual

15

3.4General Reply Format

For commands that have a reply string, the reply terminator <RT> is always transmitted and is always <CR> followed by <LF>.

<RT> is expected with all commands but only shown in some of the following sections for clarity (e.g.: Section 5.3 Diagnostics Mode).

Command Terminator

Command strings must end with the command terminator <CT>.

For Dumb Terminal applications press the carriage return (or enter) key.

With a Terminal Emulator communications application, the application must be set-up to give the required output when the carriage return (or enter) key is pressed.

For Custom applications, <CR> (decimal 13) or <CRLF> (decimal 13 followed by decimal 10) must be added to the end of each complete command string.

<CT> is expected with all commands except for some interactive commands where further explanation is necessary (e.g.: Section 4 User Calibration).

3.5Error Message

Error messages are generated from two sources:

a.The command decoder. A command error causes the return of an error message and stops the rest of the command string. The instrument sends an error message corresponding to the command error.

b.A pressure sensor fault causes an error message to be sent. Serious faults block the transmission of pressure data and sends the error message in place of data.

K177 Issue No. 3

Image 23
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Accuracy SpecificationDimensions DescriptionPressure connection Power supplyPressure measurement Data CommunicationsAccessories Pressure Connection Mounting the Instrument FigureInstrument Dimensions Communication connections Electrical ConnectionConnections for remote on/off control Changing the communication interface standard Figure CommunicationsLK1 LK2 PL3PL1 Screening RPTSystem Operation Command SummaryUser calibration Command Function MeasurementCR Crlf Command Terminators Set-upCommand Strings General Command FormatSingle commands Page General Reply Format Error MessageData out of range Auto send reading Detailed Command Definitions Measurement CommandsReturn stored reading Get readingAuto Send from switch-on Unit Set-up Set-up CommandsFilter Set-up Command F, step, averageSet Resolution Set Communications Options Identity Command I,S/PIN, Name ValueCR formatStatus User Calibration Linearity/Drift Compensation  Calibration CheckPIN PIN, New PIN Change PINGeneral LED IndicatorDiagnostic Mode Diagnostic ModeCleaning Approved Service Agents