GE RPT 301 user manual Description, Specification, Accuracy, Ranges, Dimensions

Page 9

RPT 301 User Manual

1

1 Introduction

1.1Description

The RPT 301 Digital Output Pressure Transducer measures air pressure applied to the connection port and produces an electrical output. It comprises a resonant sensor connected to a printed circuit board to produce a serial digital output and is contained in a metal enclosure to IP65 standard. The instrument is designed to communicate with compatible equipment using a serial communications system which can be selected by the user for either RS232 or RS485 standard communications interface. The power supply for the instrument is a wide tolerance DC voltage with a low power consumption A remote start-up facility enables the instrument to be remotely switched on and off to save power.

1.2Specification

Accuracy

Non-linearity, hysteresis and repeatability:

Standard

± 0.02% FS

Option A

±0.01% FS

Temperature affects:

±0.02% FS over operating temperature range

Long term stability:

 

Standard

150 ppm per 12 months

Option A

100 ppm per 12 months

Ranges

 

..................................................................

750 to 1150 mbar a

....................................................................

35 to 1300 mbar a

....................................................................

35 to 2620 mbar a

....................................................................

35 to 3500 mbar a

Dimensions

 

Weight:

0.65 kg

Size:

125 mm x 80 mm x 57 mm

Environment

-20 to + 60… C

Operating temperature range

Storage temperature range

-40 to + 80… C

Sealing:

IP65

K177 Issue No. 3

Image 9
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Dimensions SpecificationDescription AccuracyPressure measurement Power supplyData Communications Pressure connectionAccessories Pressure Connection Mounting the Instrument FigureInstrument Dimensions Communication connections Electrical ConnectionConnections for remote on/off control Changing the communication interface standard Figure CommunicationsLK1 LK2 PL3PL1 Screening RPTSystem Operation Command SummaryCR Crlf Command Terminators Command Function MeasurementSet-up User calibrationGeneral Command Format Single commandsCommand Strings Page General Reply Format Error MessageData out of range Return stored reading Detailed Command Definitions Measurement CommandsGet reading Auto send readingAuto Send from switch-on Unit Set-up Set-up CommandsCommand F, step, average Set ResolutionFilter Set-up Set Communications Options Command I,S/PIN, Name ValueCR format StatusIdentity User Calibration Linearity/Drift Compensation  Calibration CheckPIN PIN, New PIN Change PINDiagnostic Mode LED IndicatorDiagnostic Mode GeneralCleaning Approved Service Agents