The calibration routine requires a minimum of two pressures within the operating extremes (maximum of six pressures). The instrument software calculates an offset and gain correction to be applied to all subsequent readings using a statistical best straight line method.

For single point calibration, the gain factor is set to unity.

Notes

1For the pressure sensor to rapidly respond to the applied pressure, disable the filter by sending the filter command F, 0,0<CR>.

2Before sending the ‘C’ command set the instrument to operate in the pressure measurement units required for calibration. i.e. U, 0<CR> selects mbar. Always use the same units when entering applied pressures.

Command:

C, PIN

Reply:

APPLY PRESSURE nn

 

apply pressure, when steady send <CT>

Reply:

ENTER APPLIED PRESSURE nn

 

type pressure value - decimal number followed by

<CT>

 

Reply:

MORE? (Y/N)

 

type Y <CT> to calibrate more points

or

 

 

type N <CT> to complete calibration point entry

Reply (if N):

SLOPE XX INTERSECT YY

 

ACCEPT CALIBRATION (Y/N)

 

type Y <CT>

 

to accept and permanently update the calibration

or

 

 

type N <CT> to terminate without saving

Reply (if N):

CALIBRATION TERMINATED

 

end of calibration procedure

Reply (if Y):

ENTER CAL DATE

 

type date in the format DD/MM/YY

Reply:

CALIBRATION COMPLETE

 

Where:nn = pressure point 1-6

K177 Issue No. 3

Image 32
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Specification DimensionsDescription AccuracyPower supply Pressure measurementData Communications Pressure connectionAccessories Mounting the Instrument Figure Pressure ConnectionInstrument Dimensions Electrical Connection Communication connectionsConnections for remote on/off control Communications Changing the communication interface standard FigureLK1 LK2 PL3PL1 RPT ScreeningCommand Summary System OperationCommand Function Measurement CR Crlf Command TerminatorsSet-up User calibrationCommand Strings General Command FormatSingle commands Page Error Message General Reply FormatData out of range Detailed Command Definitions Measurement Commands Return stored readingGet reading Auto send readingAuto Send from switch-on Set-up Commands Unit Set-upFilter Set-up Command F, step, averageSet Resolution Set Communications Options Identity Command I,S/PIN, Name ValueCR formatStatus  Calibration Check User Calibration Linearity/Drift CompensationPIN Change PIN PIN, New PINLED Indicator Diagnostic ModeDiagnostic Mode GeneralCleaning Approved Service Agents