6

RPT 301 User Manual

2.3Electrical Connection

The electrical connector is a 10-way connector (MIL C26482) for serial, parallel and power connections.

Instrument Connector - Plug, Amphenol 62GB 12E12 - 10P

 

Cable Connector

-

Socket, Amphenol 62GB 16F12 - 10S

Communication connections

 

 

 

When configured for:

 

 

 

 

 

RS485

 

 

 

RS232

 

A

=

Power Supply -ve

A

=

Power Supply -ve

B

=

Power Supply +ve

B

=

Power Supply +ve

C

=

Chassis

 

C

=

Chassis

D

=

Data out +ve

 

D

=

RTS Input

E

=

Data out -ve

 

E

=

Data Out

F

=

Data In +ve

 

F

=

CTS Output

G

=

Data In -ve

 

G

=

Data In

H

=

Data Ground

 

H

=

Data Ground

J

=

Start-Up -ve

 

J

=

Start-Up -ve

K

=

Start-Up +ve

 

K

=

Start-Up +ve

Power supply

The instrument is internally fused with reverse polarity and overvoltage protection.

Note: The power input is transformer isolated from the logic circuits and the serial data link. The isolation voltage is 50V dc.

For continuous operation the ‘Start-up’ connections should be connected to the supply, make sure the polarity is correct and the current requirements are met.

When RS232 is used, the RTS line can be connected to the `Start-up' input and the RTS input on pin D so that the control computer can turn on the instrument.

K177 Issue No. 3

Image 14
Contents RPT General Electric Company. All rights reserved RPT Digital Output Pressure Transducer K177 K177 Issue No Safety Abbreviations and Symbols Iii K177 Issue No Description SpecificationDimensions AccuracyData Communications Power supplyPressure measurement Pressure connectionAccessories Mounting the Instrument Figure Pressure ConnectionInstrument Dimensions Electrical Connection Communication connectionsConnections for remote on/off control Communications Changing the communication interface standard FigureLK1 LK2 PL3PL1 RPT ScreeningCommand Summary System OperationSet-up Command Function MeasurementCR Crlf Command Terminators User calibrationCommand Strings General Command FormatSingle commands Page Error Message General Reply FormatData out of range Get reading Detailed Command Definitions Measurement CommandsReturn stored reading Auto send readingAuto Send from switch-on Set-up Commands Unit Set-upFilter Set-up Command F, step, averageSet Resolution Set Communications Options Identity Command I,S/PIN, Name ValueCR formatStatus  Calibration Check User Calibration Linearity/Drift CompensationPIN Change PIN PIN, New PINDiagnostic Mode LED IndicatorDiagnostic Mode GeneralCleaning Approved Service Agents