Apollo IMY instruction manual To 20 mA

Page 41

15.ENVIRONMENTAL CONDITIONS: Operating Temperature Range: 0 to 50°C Storage Temperature Range: -40 to 80°C

Operating and Storage Humidity: 85% max. relative humidity (non-condensing) from 0°C to 50°C.

Altitude: Up to 2000 meters

16.SERIAL COMMUNICATIONS (Optional):

Isolation To Signal Input Common: 500 Vrms for 1 min. Working Voltage: 50 V max., CAT I

Not isolated from all other commons.

Type: Bi-directional 20 mA current loop, 20 mA source provided on transmit loop. (Powers up to 7 units in a loop with internal current source).

Baud Rate: programmable 300 to 2400

Maximum address: 99 (Actual number in a single loop is limited by serial hardware specifications.)

Data Format: 10 bit frame, Odd parity (one start bit, 7 data bits, one odd parity bit, and one stop bit.)

Serial Hardware Specifications:

SO - Output Transistor Rating: Vmax = 30 VDC, VSAT = 1Vmax at 20 mA. Note: This will allow up to 28 units maximum in each loop.

SI - Input Diode Rating: VF = 1.25 VTYP; 1.5 VMAX

Note: The compliance voltage rating of the source must be greater than the sum of the voltage drops around the loop. (Typically a 30 VDC powered source would be capable of operating between 18 and 22 units in a loop.)

17.ALARMS (Optional):

Solid State: Two, isolated, sinking open collector NPN transistors acting in parallel with relays. Imax: 100 mA. VSAT = 1.0 V @ 100 mA. Vmax: 30 VDC.

Isolation To Signal Input Common: 500 Vrms for 1 min. Working Voltage: 50 V max., CAT I

Not isolated from all other commons.

Relays:

Isolation To Signal Input Common: 2300 Vrms for 1 min. Working Voltage: 300 V max., CAT II

Type: Form C (2)

Max. Rating: 5 Amps @ 120/240 VAC or 28 VDC (resistive load), 1/8 hp @ 120 VAC (inductive load).

Relay Life Expectancy: 100,000 cycles at max. rating. (As load level decreases, life expectancy increases.)

18.ANALOG OUTPUT (Optional): Digital scaling and offsetting Isolation To Signal Input Common: 500 Vrms for 1 min.

Working Voltage: 50 V max., CAT I

Not isolated from all other commons.

4 to 20 mA:

Accuracy: 0.1% of full scale

Resolution: 12 bits

Compliance Voltage: 10 VDC (500 ￿ max. loop resistance)

0 to 10 VDC:

Accuracy: ±(0.1% of reading + 35 mV)

Resolution: 12 bits

Min. Load Resistance: 10 K￿ (1 mA max.)

19. PEAK/VALLEY/SLOPE/OFFSET (Optional):

Peak and Valley recording. Programmable temperature offset and slope.

-39-

Image 41
Contents Apollo Intelligent Meter Series Introduction Table of Contents Page Definition of Terms Safety InformationSafety Summary Theory of Operation General DescriptionBlock Diagram Programming the IMY Programming and Operating the IMYDisplay Module #2 Program Temperature Display Offset and Slope Enter Alarm Values Display Alarm ValuesReset Latched Alarms Reset PEAK/VALLEY Memory BufferEnter Offset Value Temperature Offset ValueProgram Digital Filtering Module #4 Program Digital Filter and Remote InputProgram Function of E1-CON and Optional E2-CON Cont’d Program INTEGRATOR/TOTALIZER Time Base Module #5 Program INTEGRATOR/TOTALIZERProgram Decimal Point Position for the INTEGRATOR/TOTALIZER Module #6 Program ALARM/SETPOINT Alarm #2 High or LOW Acting Alarm #2 Assignment to Input or INTEGRATOR/ TotalizerProgram Value for Alarm #2 Unit Address Number Baud RatePrint Request Function Full or Abbreviated TransmissionAnalog Output Source Module #8 Program RE-TRANSMITTED Analog OutputAnalog Output LO Display Value Analog Output HI Display ValueEnter Access Code Module #9 Service OperationsEnter Zero Reference Apply Precision ResistanceQuick Programming Operating the IMYSlope Factory ConfigurationProgramming Example Temperature Monitoring Example PEAK/VALLEY Offset and Slope Display TemperatureINTEGRATOR/TOTALIZER INTEGRATOR/TOTALIZER SET-UP INTEGRATOR/TOTALIZER ExampleBasic SET-UP 125 x 827 Alarm Timing Diagrams Alarms OptionalCommunication Format General DescriptionCommand String Examples Sending Commands to the IMYTiming Diagrams Wiring Connections Current Loop InstallationReceiving Data from the IMY Serial Terminal DescriptionsSerial Communications Examples Connecting to AN RLC PrinterProcess Controlling System RE-TRANSMITTED Analog Output Optional To 20 mA Calibration Analog Output CalibrationSelect AC Power 115/230 VAC Appendix a Installation & ConnectionsInstallation Environment EMC Installation Guidelines Power Wiring Wiring ConnectionsOpen Thermistor Detection Display Ululul Appendix B Specifications and DimensionsTo 20 mA Certifications and Compliances Safety Problem Possible Cause Remedies Appendix C Troubleshooting GuideDecimal Point Positon Appendix D Programmable FunctionsAppendix D Programmable Functions Cont’d E2-CON Appendix E Ordering InformationLimited Warranty IMY/IM-A 3/02 Drawing NO. LP0311