Emerson SPS 4001B Calibration Gas Solenoids, Pressure Switch, Circuit Board, Pressure Regulator

Page 10

SPS 4001B

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

Calibration Gas Solenoids

The solenoids sequence the calibration gases. One solenoid controls calibration gas 1 (high calibration gas), and the other controls calibration gas 2 (low calibration gas). The solenoids activate and deactivate to allow the calibration gases to flow between the sequencer and Oxymitter 4000.

Pressure Switch

The pressure switch detects if the pressure of a calibration gas is low, which can be caused by an empty gas bottle, a disconnected gas line, etc. Calibration is prohibited when calibration gas pressure is low.

Circuit Board

The circuit board converts the incoming line voltage from AC to DC for use by the solenoids, terminations, and the programmable logic device. The power circuit is protected by a 1.6 A, 250 V, fast acting fuse.

The board contains a programmable logic device (PLD) that has the electronics to energize and deenergize the solenoids based on a signal from the Oxymitter 4000.

A terminal strip provides convenient access for all signal and power user connections.

Calibration Gas Flow Meter

The calibration gas flow meter indicates the flow rate of calibration gas flowing to the Oxymitter 4000.

Reference Air Flow Meter

The reference air flow meter indicates the amount of reference air continuously flowing to the Oxymitter 4000.

Pressure Regulator

The pressure regulator ensures the instrument air (reference air) flowing to the Oxymitter 4000 is at a constant pressure [20 psi (138 kPa)]. The regulator also has a filter to remove particulates in the reference air and a drain valve to bleed the moisture that collects in the filter bowl.

1-4

Image 10
Contents SPS 4001B Single Probe Autocalibration Sequencer Page Table of Contents TOC-2 SPS 4001B Single Probe Autocalibration Sequencer Essential InstructionsPreface SymbolsDefinitions Component Checklist Description and SpecificationsOverview Specifications SpecificationsPhysical Description ManifoldCircuit Board Pressure SwitchCalibration Gas Solenoids Calibration Gas Flow MeterTheory of Operation SPS 4001B Calibration Setup Section Installation Overview Mechanical InstallationReference Air Figure GAS ConnectionsCalibration Gas Figure Analytical Electrical ConnectionsGND SPS 4001B Overview Calibration Requirements Section OperationCalibration Gas Tanks Liquid Carbonic GAS Corp Specialty GAS Laboratories Check ValveCAL Recommended Timed IntervalSEMI-AUTOMATIC Calibration Maintenance and Service Fuse ReplacementSPS 4001B, Exploded View Installation REMOVE/INSTALL ChassisRemoval Terminal Connections Circuit Board Replacement Solenoid ReplacementPressure Switch Replacement Troubleshooting SPS 4001B TroubleshootingRelay Failure SPS 4001B Fault Finding SPS 4001B Troubleshooting Flowchart Sheet 1 SPS 4001B Troubleshooting Flowchart Sheet 2 SPS 4001B Replacement Parts SPS 4001B Calibration Components Index No Part Number DescriptionSPS 4001B Appendix a Safety Data Safety Instructionsimportant Belangrijk Vigtigt SPS 4001B Tärkeää SPS 4001B Wichtig Importante Viktig SPS 4001B SPS 4001B Viktigt SPS 4001B Appendix B Return of Materials Returning MaterialSPS 4001B Index Index-2 Warranty IM-106-340AC-1, Rev .0 April SPS 4001B Serial no Order no