Emerson SPS 4001B instruction manual Section Installation, Overview Mechanical Installation

Page 13

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

SPS 4001B

Section 2

Installation

OVERVIEW

MECHANICAL INSTALLATION

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1 Mechanical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1 Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-2 Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4

This section describes SPS 4001B installation.

Before starting to install this equipment, read the "Safety instructions for the wiring and installation of this apparatus" in Section A: Safety Data. Failure to follow the safety instructions could result in serious injury or death.

Secure equipment cover and ground leads after installation. Failure to secure covers and ground leads could result in serious injury or death.

The outline drawing in Figure 2-1shows mounting centers and clearances of the SPS 4001B. The unit is designed to mount on a wall, bulkhead, or pipe. Ensure the unit is installed according to the following specifications.

1.Install the unit no further than 300 ft (91 m) from the Oxymitter 4000 and no further than 1000 ft (303 m) from the electronics package or any customer-supplied remote input or relay output in the control room.

2.Locate the unit where the ambient temperature is between -25° and 149°F (-5° and 65°C) or -40° and 149°F (-40° and 65°C) with optional heater.

http://www.processanalytic.com

Image 13
Contents SPS 4001B Single Probe Autocalibration Sequencer Page Table of Contents TOC-2 Essential Instructions SPS 4001B Single Probe Autocalibration SequencerPreface SymbolsDefinitions Component Checklist Description and SpecificationsOverview Specifications SpecificationsManifold Physical DescriptionCalibration Gas Solenoids Pressure SwitchCircuit Board Calibration Gas Flow MeterTheory of Operation SPS 4001B Calibration Setup Overview Mechanical Installation Section InstallationReference Air Figure GAS ConnectionsCalibration Gas Figure Analytical Connections ElectricalGND SPS 4001B Overview Calibration Requirements Section OperationCalibration Gas Tanks Check Valve Liquid Carbonic GAS Corp Specialty GAS LaboratoriesTimed Interval CAL RecommendedSEMI-AUTOMATIC Calibration Fuse Replacement Maintenance and ServiceSPS 4001B, Exploded View Installation REMOVE/INSTALL ChassisRemoval Terminal Connections Solenoid Replacement Circuit Board ReplacementPressure Switch Replacement SPS 4001B Troubleshooting TroubleshootingRelay Failure SPS 4001B Fault Finding SPS 4001B Troubleshooting Flowchart Sheet 1 SPS 4001B Troubleshooting Flowchart Sheet 2 SPS 4001B Replacement Parts SPS 4001B Index No Part Number Description Calibration ComponentsSPS 4001B Appendix a Safety Data Safety Instructionsimportant Belangrijk Vigtigt SPS 4001B Tärkeää SPS 4001B Wichtig Importante Viktig SPS 4001B SPS 4001B Viktigt SPS 4001B Returning Material Appendix B Return of MaterialsSPS 4001B Index Index-2 Warranty IM-106-340AC-1, Rev .0 April SPS 4001B Serial no Order no