Instruction Manual
SPS 4001B
Section 3 | Operation |
OVERVIEW
CALIBRATION REQUIREMENTS
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page
This section specifies the requirements to set up an Oxymitter 4000 calibration and how to verify the calibration gas flow setup. It also explains the differences between automatic and
Calibration Gas Tanks
Two tanks of precision calibration gas mixtures are required. Recommended calibration gases are nominally 0.4% and 8.0% oxygen in nitrogen.
Do not use 100% nitrogen as a low gas (zero gas). It is suggested that gas for the low (zero) be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon concentrations of more than 40 parts per million. Failure to use proper gases will result in erroneous readings.