Emerson SPS 4001B instruction manual Physical Description, Manifold

Page 9

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

SPS 4001B

Table 1-1. Calibration

Components

PART NUMBER

DESCRIPTION

 

1A99119G01

Two disposable calibration gas bottles - 0.4% and 8% O2, balance

 

 

nitrogen - 550 liters each (1) (2)

 

1A99119G02

Two flow regulators for calibration gas bottles

 

1A99119G03

Bottle rack

NOTES

(1)Calibration gas bottles cannot be shipped via airfreight.

(2)When the bottles are used with "CALIBRATION RECOMMENDED" features, the bottles should provide 2 to 3 years of calibrations in normal service.

PHYSICAL DESCRIPTION

The main components of the SPS 4001B are illustrated in Figure 1-2.

Manifold

The manifold provides porting to and from the calibration solenoids. The manifold ports are accessible from outside the enclosure for making piping connections. The Oxymitter 4000 calibration gases are piped into and sequenced through the calibration gas solenoids mounted on the manifold.

Figure 1-2. SPS 4001B Components

CALIBRATION GAS

FLOW METER

REFERENCE AIR

FLOW METER

CIRCUIT

BOARD

REFERENCE

AIR PRESSURE

REGULATOR

PRESSURE

SWITCH

CALIBRATION GAS 1

(LOW CAL GAS)

SOLENOID

 

SOLENOID

37740002

 

CALIBRATION GAS 2

 

MANIFOLD

(HIGH CAL GAS)

 

1-3

Image 9
Contents SPS 4001B Single Probe Autocalibration Sequencer Page Table of Contents TOC-2 Essential Instructions SPS 4001B Single Probe Autocalibration SequencerSymbols PrefaceDefinitions Description and Specifications Component ChecklistOverview Specifications SpecificationsManifold Physical DescriptionCalibration Gas Solenoids Pressure SwitchCircuit Board Calibration Gas Flow MeterTheory of Operation SPS 4001B Calibration Setup Overview Mechanical Installation Section InstallationGAS Connections Reference Air FigureCalibration Gas Figure Analytical Connections ElectricalGND SPS 4001B Section Operation Overview Calibration RequirementsCalibration Gas Tanks Check Valve Liquid Carbonic GAS Corp Specialty GAS LaboratoriesTimed Interval CAL RecommendedSEMI-AUTOMATIC Calibration Fuse Replacement Maintenance and ServiceSPS 4001B, Exploded View REMOVE/INSTALL Chassis InstallationRemoval Terminal Connections Solenoid Replacement Circuit Board ReplacementPressure Switch Replacement SPS 4001B Troubleshooting TroubleshootingRelay Failure SPS 4001B Fault Finding SPS 4001B Troubleshooting Flowchart Sheet 1 SPS 4001B Troubleshooting Flowchart Sheet 2 SPS 4001B Replacement Parts SPS 4001B Index No Part Number Description Calibration ComponentsSPS 4001B Appendix a Safety Data Safety Instructionsimportant Belangrijk Vigtigt SPS 4001B Tärkeää SPS 4001B Wichtig Importante Viktig SPS 4001B SPS 4001B Viktigt SPS 4001B Returning Material Appendix B Return of MaterialsSPS 4001B Index Index-2 Warranty IM-106-340AC-1, Rev .0 April SPS 4001B Serial no Order no