Emerson instruction manual SPS 4001B Troubleshooting

Page 29

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

SPS 4001B

Section 5 Troubleshooting

OVERVIEW

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 5-1 SPS 4001B Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . page 5-1

This section describes the SPS 4001B troubleshooting procedures. Additional troubleshooting information can be found in the Oxymitter 4000 Oxygen Transmitter Instruction Manual.

Secure equipment cover and ground leads after troubleshooting. Failure to secure covers and ground leads could result in serious injury or death.

SPS 4001B TROUBLESHOOTING

Use the CAL FAIL and IN CAL relay outputs to identify possible SPS faults.

CAL FAIL

If a calibration was not successfully completed, the SPS 4001B sends a CAL FAIL contact indication to the control room. To determine if the SPS 4001B caused the failed calibration, go to the Oxymitter 4000 site to view the keypad. Or, access the HART/AMS menus. For more information on HART/AMS, refer to the HART/AMS section in the Oxymitter 4000 Oxygen Transmitter Instruction Manual.

If no alarms are indicated on the keypad, LOI, or in the HART/AMS STATUS submenu, the calibration did not fail because of an Oxymitter 4000 fault. Therefore, a calibration gas flow problem occurred. Refer to Table 5-1or Figure 5-1to troubleshoot the SPS 4001B.

If the LAST CAL FAILED alarm is indicated on the keypad, LOI, or in the HART/AMS STATUS submenu, the failure is due to either a bad Oxymitter 4000 cell or a calibration gas flow problem.

http://www.raihome.com

Image 29
Contents SPS 4001B Single Probe Autocalibration Sequencer Page Table of Contents TOC-2 Essential Instructions SPS 4001B Single Probe Autocalibration SequencerDefinitions SymbolsPreface Overview Description and SpecificationsComponent Checklist Specifications SpecificationsManifold Physical DescriptionCalibration Gas Solenoids Pressure SwitchCircuit Board Calibration Gas Flow MeterTheory of Operation SPS 4001B Calibration Setup Overview Mechanical Installation Section InstallationCalibration Gas Figure GAS ConnectionsReference Air Figure Analytical Connections ElectricalGND SPS 4001B Calibration Gas Tanks Section OperationOverview Calibration Requirements Check Valve Liquid Carbonic GAS Corp Specialty GAS LaboratoriesTimed Interval CAL RecommendedSEMI-AUTOMATIC Calibration Fuse Replacement Maintenance and ServiceSPS 4001B, Exploded View Removal REMOVE/INSTALL ChassisInstallation Terminal Connections Solenoid Replacement Circuit Board ReplacementPressure Switch Replacement SPS 4001B Troubleshooting TroubleshootingRelay Failure SPS 4001B Fault Finding SPS 4001B Troubleshooting Flowchart Sheet 1 SPS 4001B Troubleshooting Flowchart Sheet 2 SPS 4001B Replacement Parts SPS 4001B Index No Part Number Description Calibration ComponentsSPS 4001B Appendix a Safety Data Safety Instructionsimportant Belangrijk Vigtigt SPS 4001B Tärkeää SPS 4001B Wichtig Importante Viktig SPS 4001B SPS 4001B Viktigt SPS 4001B Returning Material Appendix B Return of MaterialsSPS 4001B Index Index-2 Warranty IM-106-340AC-1, Rev .0 April SPS 4001B Serial no Order no