SPS 4001B
Instruction Manual
Figure 1-3. SPS 4001B Calibration Setup
OXYMITTER 4000 (BOTTOM VIEW)
OXYMITTER 4000
LOGIC I/O
CHECK VALVE (SEE NOTE 1)
GAS |
|
|
CALIBRATION |
| INSTRUMENT |
|
| |
| SPS | AIR IN |
| (SEE NOTE 2) | |
| 4001B | |
|
|
NOTES: |
| |
| 1. A CHECK VALVE IS REQUIRED AT THE | |
| OXYMITTER 4000 (BETWEEN THE | |
| CALIBRATION GAS FITTING AND THE | |
| GAS LINE) TO PREVENT THE MIGRATION | |
| OF PROCESS GASES DOWN THE | |
| CALIBRATION GAS LINE. |
|
| 2. CLEAN, DRY INSTRUMENT AIR IS | |
| RECOMMENDED FOR REFERENCE AIR. | |
| NO REFERENCE AIR IS REQUIRED IF | |
2) | AMBIENT AIR CONDITIONS CONTAIN | |
20.95% OXYGEN. |
| |
(SEENOTE |
| |
|
| |
AIR |
|
|
REFERENCE |
| 37740005 |
| CALIBRATION | CALIBRATION |
| GAS 1 | GAS 2 |
| (HIGH O2) | (LOW O2) |