Emerson SPS 4001B instruction manual GAS Connections, Reference Air Figure, Calibration Gas Figure

Page 14

SPS 4001B

Instruction Manual

IM-106-340AC-1, Rev. 1.0 April 2005

GAS CONNECTIONS

Use the following procedures to connect the reference air and the calibration gases.

Reference Air (Figure 2-1)

The Oxymitter 4000 requires a source of reference air that is constantly 20.95% O2. Clean and dry instrument air provides the most stable source of reference air. If instrument air is unavailable, the reference air port of the probe can be left open, permitting ambient air to passively diffuse into the probe.

1.Connect the instrument air supply to the INST AIR IN port of the manifold.

2.The pressure regulator is factory set at 20 psi (138 kPa). If necessary, readjust by turning the knob on the top of the regulator until the desired pressure is obtained.

3.Next, connect the reference air from the REF GAS OUT port of the manifold to the REF GAS port on the Oxymitter 4000.

Calibration Gas (Figure 2-1)

1.Connect O2 calibration gas 1 (high calibration gas) to the HI GAS IN port on the manifold. Ensure the calibration gas pressure is set at 20 psi (138 kPa).

Instrument air is not recommended for the high calibration gas. Do not use 100% nitrogen as a low gas (zero gas). It is suggested that the low (zero) gas be between 0.4% and 2.0% O2. Do not use gases with hydrocarbon concentrations of more than 40 parts per million.

Failure to use proper gases will result in erroneous readings.

2.Connect O2 calibration gas 2 (low calibration gas) to the LOW GAS IN port on the manifold. Ensure the calibration gas pressure is set at 20 psi (138 kPa).

3.Connect the calibration gas from the CAL GAS OUT manifold fitting to the check valve connected to the CAL GAS port on the Oxymitter 4000.

2-2

Image 14
Contents SPS 4001B Single Probe Autocalibration Sequencer Page Table of Contents TOC-2 SPS 4001B Single Probe Autocalibration Sequencer Essential InstructionsDefinitions SymbolsPreface Overview Description and SpecificationsComponent Checklist Specifications SpecificationsPhysical Description ManifoldCircuit Board Pressure SwitchCalibration Gas Solenoids Calibration Gas Flow MeterTheory of Operation SPS 4001B Calibration Setup Section Installation Overview Mechanical InstallationCalibration Gas Figure GAS ConnectionsReference Air Figure Analytical Electrical ConnectionsGND SPS 4001B Calibration Gas Tanks Section OperationOverview Calibration Requirements Liquid Carbonic GAS Corp Specialty GAS Laboratories Check ValveCAL Recommended Timed IntervalSEMI-AUTOMATIC Calibration Maintenance and Service Fuse ReplacementSPS 4001B, Exploded View Removal REMOVE/INSTALL ChassisInstallation Terminal Connections Circuit Board Replacement Solenoid ReplacementPressure Switch Replacement Troubleshooting SPS 4001B TroubleshootingRelay Failure SPS 4001B Fault Finding SPS 4001B Troubleshooting Flowchart Sheet 1 SPS 4001B Troubleshooting Flowchart Sheet 2 SPS 4001B Replacement Parts SPS 4001B Calibration Components Index No Part Number DescriptionSPS 4001B Appendix a Safety Data Safety Instructionsimportant Belangrijk Vigtigt SPS 4001B Tärkeää SPS 4001B Wichtig Importante Viktig SPS 4001B SPS 4001B Viktigt SPS 4001B Appendix B Return of Materials Returning MaterialSPS 4001B Index Index-2 Warranty IM-106-340AC-1, Rev .0 April SPS 4001B Serial no Order no