Sierra SL-800 instruction manual Setup for Piston or Bell Provers

Page 12

7.0Installation Diagrams and Application Guide

7.1Comparison of Cal=Trak with Piston or Bell Provers

Piston or bell provers have a much longer measurement time than the Cal=Trak. For this reason, it is possible to compare them simultaneously, but certain precautions must be observed. When the Cal=Trak begins its cycle, the piston’s weight causes the internal pressure to rise by about 0.001 atmospheres (~0.1 kPa). If a simple pressure regulator feeds the test chain, we are simply using the resistance of the entire flow chain to set our flow rate. The rate will then change significantly when the Cal=Trak is in its measurement cycle. This will cause the actual flow measured during the Cal=Trak cycles to be less than the average flow seen by the piston or bell prover.

To render this effect insignificant, the flow must not be affected significantly by the Cal=Trak’s cyclic pressure increase. This can be achieved by use of a sonic nozzle as the stable flow source, or by feeding a fixed restrictor with a precisely regulated pressure of more than 200 kPa, as is used in factory calibration of the Cal=Trak. Note that at 200 kPa (30 PSI), the dynamic flow decrease of a simple restrictor caused by the piston’s weight will be about 0.05%. In certain circumstances, the sonic nozzle or porous plug flow generator may be replaced with a mass flow controller (MFC) specifically tailored to this task. An example is the Sierra Model C100L-10, -24 and -44, each specifically built to provide a stable flow source for the Cal=Trak calibrator.

For this type of calibration, we can use the setup shown in Figure 5. The adjustable regulator is used to set the flow rate within the range of a properly sized flow restrictor. A piston or bell prover cycle is instituted. The Cal=Trak and the prover can then be alternately measured using the fixed flow source.

Gas

Fixed

Adjustable

Sonic

Supply

Regulator

Regulator

Nozzle

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Stable Pressure

OR Porous Plug

 

> 200 kPa

 

Cal=TrakDryCal

Piston or

Bell Prover

Figure 5 Setup for Piston or Bell Provers

An alternative approach can be used with piston provers, as shown in Figure 6. A cycle is initiated on the prover, which is much slower than a Cal=Trak. The Cal=Trak is then started in a cyclical mode, averaging its flow. Before the prover ends its cycle, the Cal=Trak is stopped and the average flow read.

The Cal=Trak can be set for sufficient cycles in its average to allow interruption by the “stop” button, or smaller averages, such as 5 or 10 readings, can be taken during the prover cycle. It should be noted that the periodic pressure pulses might cause oscillations in bell provers, reducing the bell prover’s accuracy somewhat.

12

Image 12
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page General Description Theory of OperationIdealized Automatic Piston Prover Cal=Trak Layout Unpacking Checklist Your Cal=Trak SL-800 Base IncludesEach of Your Cal=Trak SL-800 Flow Cells Includes Cal=Trak Installation Attaching & Removing Flow Cells Connecting the Cal=Trak to a Flow Source Cal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Application Precautions Inventory Dead VolumeComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Alternative Setup for Piston Provers Vacuum Comparison of Cal=Trak with Piston or Bell ProversSetup for LFEs Comparison of Cal=Trak with Sonic Nozzle Transfer StandardsVacuum Setup for Sonic Nozzle Transfer Standard Calibration of Mass Flow Controllers MFCsCalibration of Mass Flow Meters MFMs Calibration of Rotameters Variable Area Flow MetersCharging the Cal=Trak Battery Operating InstructionsTurning the Cal=Trak On Turning the Cal=Trak OffCal=Trak Keypad How To Use the Cal=Trak Keypad Low Battery IndicatorGeneral Menu Navigation Factory Default Settings Parameters Factory Settings Optional SettingsTaking Readings Setting User Preferences Option 1, Sensor FactorSetup Menu 2, Reading Type, # in Average & Minutes/Reading Setup Menu 4, Date, Time & Battery VoltageSetup Menu 5, Date & Time Formats Option 1, DateOption 2, Time Option 3, Auto OffBattery System Battery Maintenance & StorageMaintenance Quality Assurance Leak Test Procedure To Initiate the Leak TestCalibration Leak Test Illustration Leak Test in ProgressReturning Your Unit for Calibration or Service Shipment Replacement Parts & AccessoriesCalibration Includes Part DescriptionAdditional Information Limited Warranty Charge LED does not light when charger plugged Settings have been lost or seem to have changedPiston Error appears on display Unit suddenly turns offPiston does not move when the Read or Auto is pushed When taking readings, the flow seems offDuring Leak Test, Leakage 9.999 ccm appears on display Resetting the Cal=Trak Contact Sierra for AssistanceModel Optimum Flow Range Flow Cell Standardized 15-30CInstallation Operating Instructions for Cal=Soft Software IntroductionRequirements Worksheet descriptionsSetup Cal=Trak IntroductionSetup DUT Setup Cal=SoftPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Software Troubleshooting Serial Port Selection IssueSerial Port Configuration Issue