Appendix B: Cal=Trak Specifications

Configuration: Interchangeable modular flow cells, all calibrated elements except crystal time base are in cell

Base Unit

Size 6" W x 10.5" D x 4" H ￿ 15.2 cm W x 26.7 cm D x 10.2 cm H

Weight 4.2 lbs ￿ 1.9 kg

Flow Cells

Size 13" H x 4” diameter ￿ 33 cm H x 10.2 cm diameter

Weight 5.8 lbs ￿ 2.6 kg

Flow Ranges:

 

Model

Optimum Flow Range

Cal=Trak SL-800-10

5.0 sccm–500 sccm

Cal=Trak SL-800-24

50 sccm–5 slm

Cal=Trak SL-800-44

500 sccm–50 slm￿

￿Maximum flow rate is 50 slm at sea level, at higher altitudes piston may not return at flow rates higher than 40 slm.

Absolute Accuracy, Single Readings

All accuracies are based on % of readings. Averaging multiple readings will increase repeatability and accuracy.

Flow Cell

Standardized (15°–30°C)

Cal=Trak SL-800-10

±0.15%

Cal=Trak SL-800-24

±0.15%

Cal=Trak SL-800-44

±0.15%

Suitable Gases Non-corrosive, humidity less than 70%, non-condensingTime per Reading Approximately 1 to 60 seconds, flow & cell dependent Operating Modes Single cycle, continuous cycling or 1 to 255 minute intervals Measuring Cell Graphite composite piston in borosilicate glass cylinder Temperature Range 15–30 C

Temperature & Pressure Sensors Located below piston in flow cell; barometric pressure sensor located in base unit Valve magnetic acting design (to prevent heating of gas stream), self-relieving

Battery System Internal continuously chargeable sealed 6V lead-acid battery AC Charger Adapter 100–240 VAC 47–63 Hz

Self-TestPiston Tare Value (Leak Rate)

Outputs Illuminated LCD display, RS-232 serial port (CSV format) Warranty 1 year

Note: Annual calibration offered by Sierra is elective and is not included as a warranty item. Software Revision 2.03

All specifications are subject to change. Please contact Sierra or visit our web site at www.sierrainstruments.com for the most current product information.

32

Image 32
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page General Description Theory of OperationIdealized Automatic Piston Prover Cal=Trak Layout Unpacking Checklist Your Cal=Trak SL-800 Base IncludesEach of Your Cal=Trak SL-800 Flow Cells Includes Cal=Trak Installation Attaching & Removing Flow Cells Connecting the Cal=Trak to a Flow SourceCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Application Precautions Inventory Dead VolumeComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Alternative Setup for Piston Provers Vacuum Comparison of Cal=Trak with Piston or Bell ProversSetup for LFEs Comparison of Cal=Trak with Sonic Nozzle Transfer StandardsVacuum Setup for Sonic Nozzle Transfer Standard Calibration of Mass Flow Controllers MFCsCalibration of Mass Flow Meters MFMs Calibration of Rotameters Variable Area Flow MetersCharging the Cal=Trak Battery Operating InstructionsTurning the Cal=Trak On Turning the Cal=Trak OffGeneral Menu Navigation Cal=Trak Keypad How To Use the Cal=Trak KeypadLow Battery Indicator Taking Readings Factory Default SettingsParameters Factory Settings Optional Settings Setting User Preferences Option 1, Sensor FactorSetup Menu 2, Reading Type, # in Average & Minutes/Reading Setup Menu 4, Date, Time & Battery VoltageSetup Menu 5, Date & Time Formats Option 1, DateOption 2, Time Option 3, Auto OffMaintenance Battery SystemBattery Maintenance & Storage Quality Assurance Leak Test Procedure To Initiate the Leak TestCalibration Leak Test Illustration Leak Test in ProgressReturning Your Unit for Calibration or Service Shipment Replacement Parts & AccessoriesCalibration Includes Part DescriptionAdditional Information Limited Warranty Charge LED does not light when charger plugged Settings have been lost or seem to have changed Piston Error appears on display Unit suddenly turns offDuring Leak Test, Leakage 9.999 ccm appears on display Piston does not move when the Read or Auto is pushedWhen taking readings, the flow seems off Resetting the Cal=Trak Contact Sierra for AssistanceModel Optimum Flow Range Flow Cell Standardized 15-30CInstallation Operating Instructions for Cal=Soft Software IntroductionRequirements Worksheet descriptionsSetup Cal=Trak IntroductionSetup DUT Setup Cal=SoftPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Serial Port Configuration Issue Software TroubleshootingSerial Port Selection Issue