Press the Read button to initiate a single reading.

Press the Auto button to initiate multiple readings.

Press the Stop button to stop current flow reading and open valve.

Press the Reset button to clear the display of current data.

8.5Setting User Preferences

The Sierra Cal=Trak offers enhanced electronics options that allow the user to define parameters specific to their application. There are six Setup Menus in all. For a comprehensive flowchart of the complete menu tree, please see Appendix B.

General Menu Navigation

Use the Up/Down arrows to navigate between different lines within a menu. Use the Enter button to select a field to be modified and to lock in any changes. Use the keypad for data entry such as # of readings in an averaging sequence; use the Up/Down arrows to toggle fields such as temperature or pressure units. Use the Fwd and Back keys to advance to the following menu or return to the previous menu.

Keypad Function

The keypad is used for both numeric and alpha character entry. When pressed, each alphanumeric button will display its corresponding number. If pressed again within two seconds the unit will display the first corresponding letter for that key. Repeat this process for additional letters on the same key. A space may be inserted by pressing the 0 key twice. To go back, use the back arrow. The Cal=Trak will overwrite previous entries. Use the Enter key to lock in any selections.

8.6Setup Menu 1, Calibration ID #, Gas Constant, Calibration Type

Run

Main Menu

Setup

Leak Test

To enter “Setup Menu – 1” from the “Main Menu”, select Setup and press Enter.

Setup Menu – 1

Sensor Factor

1.000

Cal. Type

Std.

Option 1, “Sensor Factor”

This option changes the gas constant or K-factor for instruments that are used with any gas other than air or nitrogen, but originally calibrated by the device manufacturer using air or nitrogen and a factor. Gas conversion factors vary by device manufacturer. Please consult the instrument manufacturer’s gas conversion multipliers for the device under test and enter the appropriate factor here. The input range for this option is 0.200–2.000. The default setting is for 1.000, the setting for air or nitrogen.

20

Image 20
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page General Description Theory of OperationIdealized Automatic Piston Prover Cal=Trak Layout Unpacking Checklist Your Cal=Trak SL-800 Base IncludesEach of Your Cal=Trak SL-800 Flow Cells Includes Cal=Trak Installation Attaching & Removing Flow Cells Connecting the Cal=Trak to a Flow SourceCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Application Precautions Inventory Dead VolumeComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Alternative Setup for Piston Provers Vacuum Comparison of Cal=Trak with Piston or Bell ProversSetup for LFEs Comparison of Cal=Trak with Sonic Nozzle Transfer StandardsVacuum Setup for Sonic Nozzle Transfer Standard Calibration of Mass Flow Controllers MFCsCalibration of Mass Flow Meters MFMs Calibration of Rotameters Variable Area Flow MetersCharging the Cal=Trak Battery Operating InstructionsTurning the Cal=Trak On Turning the Cal=Trak OffGeneral Menu Navigation Cal=Trak Keypad How To Use the Cal=Trak KeypadLow Battery Indicator Taking Readings Factory Default SettingsParameters Factory Settings Optional Settings Setting User Preferences Option 1, Sensor FactorSetup Menu 2, Reading Type, # in Average & Minutes/Reading Setup Menu 4, Date, Time & Battery VoltageSetup Menu 5, Date & Time Formats Option 1, DateOption 2, Time Option 3, Auto OffMaintenance Battery SystemBattery Maintenance & Storage Quality Assurance Leak Test Procedure To Initiate the Leak TestCalibration Leak Test Illustration Leak Test in ProgressReturning Your Unit for Calibration or Service Shipment Replacement Parts & AccessoriesCalibration Includes Part DescriptionAdditional Information Limited Warranty Charge LED does not light when charger plugged Settings have been lost or seem to have changedPiston Error appears on display Unit suddenly turns offDuring Leak Test, Leakage 9.999 ccm appears on display Piston does not move when the Read or Auto is pushedWhen taking readings, the flow seems off Resetting the Cal=Trak Contact Sierra for AssistanceModel Optimum Flow Range Flow Cell Standardized 15-30CInstallation Operating Instructions for Cal=Soft Software IntroductionRequirements Worksheet descriptionsSetup Cal=Trak IntroductionSetup DUT Setup Cal=SoftPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Serial Port Configuration Issue Software TroubleshootingSerial Port Selection Issue