Sierra SL-800 instruction manual Calibration, Leak Test Illustration Leak Test in Progress

Page 25

8B If the unit fails the leak test, the display will appear similar to the following picture. Please make a note of the number that is displayed and contact Sierra customer service.

-

9Repeat the procedure with the leak test port cap over the opposite port fitting.

10.To stop the leak test at any time press the RESET button.

Leak Test Illustration: Leak Test in Progress

Inverted Cell

Leak Test Plug

Leak Test Cable

Swagelock Fitting

Base

11.2Calibration

As a quality assurance measure, Sierra recommends annual calibration of all measurement instruments. However, how often you have your Cal=Trak calibrated is an internal quality control decision. The determining factors are whether the unit passes the internal leak test, quality system requirements if applicable, and the conditions in which the unit is used. Units used in a laboratory setting may require calibration less frequently than a unit that is used in a manufacturing environment. The annual calibration program is an elective and is therefore not included as a warranty item. Expedited “RUSH” turnaround service is also available at an additional cost. Please contact the factory for more information on available calibration services and pricing.

25

Image 25
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page Theory of Operation General DescriptionIdealized Automatic Piston Prover Your Cal=Trak SL-800 Base Includes Cal=Trak Layout Unpacking ChecklistEach of Your Cal=Trak SL-800 Flow Cells Includes Connecting the Cal=Trak to a Flow Source Cal=Trak Installation Attaching & Removing Flow CellsCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Inventory Dead Volume Application PrecautionsComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Vacuum Comparison of Cal=Trak with Piston or Bell Provers Alternative Setup for Piston ProversComparison of Cal=Trak with Sonic Nozzle Transfer Standards Setup for LFEsCalibration of Mass Flow Controllers MFCs Vacuum Setup for Sonic Nozzle Transfer StandardCalibration of Rotameters Variable Area Flow Meters Calibration of Mass Flow Meters MFMsOperating Instructions Charging the Cal=Trak BatteryTurning the Cal=Trak On Turning the Cal=Trak OffLow Battery Indicator Cal=Trak Keypad How To Use the Cal=Trak KeypadGeneral Menu Navigation Parameters Factory Settings Optional Settings Factory Default SettingsTaking Readings Option 1, Sensor Factor Setting User PreferencesSetup Menu 4, Date, Time & Battery Voltage Setup Menu 2, Reading Type, # in Average & Minutes/ReadingOption 1, Date Setup Menu 5, Date & Time FormatsOption 2, Time Option 3, Auto OffBattery Maintenance & Storage Battery SystemMaintenance To Initiate the Leak Test Quality Assurance Leak Test ProcedureLeak Test Illustration Leak Test in Progress CalibrationReplacement Parts & Accessories Returning Your Unit for Calibration or Service ShipmentCalibration Includes Part DescriptionAdditional Information Limited Warranty Settings have been lost or seem to have changed Charge LED does not light when charger pluggedPiston Error appears on display Unit suddenly turns offWhen taking readings, the flow seems off Piston does not move when the Read or Auto is pushedDuring Leak Test, Leakage 9.999 ccm appears on display Contact Sierra for Assistance Resetting the Cal=TrakFlow Cell Standardized 15-30C Model Optimum Flow RangeIntroduction Installation Operating Instructions for Cal=Soft SoftwareRequirements Worksheet descriptionsIntroduction Setup Cal=TrakSetup Cal=Soft Setup DUTPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Serial Port Selection Issue Software TroubleshootingSerial Port Configuration Issue