Sierra SL-800 Operating Instructions, Charging the Cal=Trak Battery, Turning the Cal=Trak On

Page 17

Variable Area Meter (Rotameter)

Fixed

Adjustable

Sonic

A

Regulator

Regulator

Nozzle

B

Stable Pressure

OR Porous Plug

 

> 200 kPa

 

DryCal

Cal=Trak

Figure 13 Setup for Calibrating Rotameters (Variable Area Flow Meters)

8.0Operating Instructions

Charging the Cal=Trak Battery

The Sierra Cal=Trak is powered by an internal lead-acid battery. The battery will power the instrument for 6–8 hours of continuous use and has a typical service life of approximately 2–4 years, depending on use. The Cal=Trak provides a convenient 65-minute automatic shut-off to extend battery life.

Before using your calibrator, be sure that the battery system has been adequately charged to ensure that it will perform to its specifications and maintain operation for the calibration period. If "B" is indicated in the upper right hand corner of the display during operation, recharging is required.

To charge, connect the Sierra supplied charger into a standard wall outlet. Be sure to use only Sierra-approved batteries and AC adapters/chargers for all Cal=Trak flow calibrators. Insert the Cal=Trak charger’s barrel plug end into the charging jack located on the left rear side of the SL-800 base. The green Charge LED will illuminate. Full charge takes 8 –12 hours, and the Cal=Trak can charge while being used. Upon full charge, the charger will taper to a trickle mode automatically. The unit may be charged indefinitely without damage to the battery.

Turning the Cal=Trak On

Press On to start the Cal=Trak. An opening screen will appear indicating the instrument’s software revision level followed by the “Main Menu”.

Turning the Cal=Trak Off

The Cal=Trak has a battery saving automatic shut off system. After 65 minutes of inactivity, the Cal=Trak will shut off. Alternatively, the unit can be shut off manually by pressing the Reset button followed by the number 0 from the “Main Menu”.

LCD Backlight

The Cal=Trak’s LCD display includes a backlight function to illuminate the display. The default setting for the backlight is always on. If you wish to conserve battery power you may wish to turn the backlight off or enable the backlight only when inputting information on the keypad.

To turn the backlight off, from the “Main Menu” press the Light button one time. To enable the backlight only when inputting information on the keypad press the Light button a second time. To return to the default setting of always on, press the Light button a third time.

17

Image 17
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page Theory of Operation General DescriptionIdealized Automatic Piston Prover Your Cal=Trak SL-800 Base Includes Cal=Trak Layout Unpacking ChecklistEach of Your Cal=Trak SL-800 Flow Cells Includes Connecting the Cal=Trak to a Flow Source Cal=Trak Installation Attaching & Removing Flow CellsCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Inventory Dead Volume Application PrecautionsComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Vacuum Comparison of Cal=Trak with Piston or Bell Provers Alternative Setup for Piston ProversComparison of Cal=Trak with Sonic Nozzle Transfer Standards Setup for LFEsCalibration of Mass Flow Controllers MFCs Vacuum Setup for Sonic Nozzle Transfer StandardCalibration of Rotameters Variable Area Flow Meters Calibration of Mass Flow Meters MFMsOperating Instructions Charging the Cal=Trak BatteryTurning the Cal=Trak On Turning the Cal=Trak OffGeneral Menu Navigation Cal=Trak Keypad How To Use the Cal=Trak KeypadLow Battery Indicator Taking Readings Factory Default SettingsParameters Factory Settings Optional Settings Option 1, Sensor Factor Setting User PreferencesSetup Menu 4, Date, Time & Battery Voltage Setup Menu 2, Reading Type, # in Average & Minutes/ReadingOption 1, Date Setup Menu 5, Date & Time FormatsOption 2, Time Option 3, Auto OffMaintenance Battery SystemBattery Maintenance & Storage To Initiate the Leak Test Quality Assurance Leak Test ProcedureLeak Test Illustration Leak Test in Progress CalibrationReplacement Parts & Accessories Returning Your Unit for Calibration or Service ShipmentCalibration Includes Part DescriptionAdditional Information Limited Warranty Settings have been lost or seem to have changed Charge LED does not light when charger pluggedPiston Error appears on display Unit suddenly turns offDuring Leak Test, Leakage 9.999 ccm appears on display Piston does not move when the Read or Auto is pushedWhen taking readings, the flow seems off Contact Sierra for Assistance Resetting the Cal=TrakFlow Cell Standardized 15-30C Model Optimum Flow RangeIntroduction Installation Operating Instructions for Cal=Soft SoftwareRequirements Worksheet descriptionsIntroduction Setup Cal=TrakSetup Cal=Soft Setup DUTPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Serial Port Configuration Issue Software TroubleshootingSerial Port Selection Issue