Variable Area Meter (Rotameter)
Fixed | Adjustable | Sonic | A |
Regulator | Regulator | Nozzle |
B
Stable Pressure | OR Porous Plug |
| |
> 200 kPa |
|
DryCal
Cal=Trak
Figure 13 Setup for Calibrating Rotameters (Variable Area Flow Meters)
8.0Operating Instructions
Charging the Cal=Trak Battery
The Sierra Cal=Trak is powered by an internal
Before using your calibrator, be sure that the battery system has been adequately charged to ensure that it will perform to its specifications and maintain operation for the calibration period. If "B" is indicated in the upper right hand corner of the display during operation, recharging is required.
To charge, connect the Sierra supplied charger into a standard wall outlet. Be sure to use only
Turning the Cal=Trak On
Press On to start the Cal=Trak. An opening screen will appear indicating the instrument’s software revision level followed by the “Main Menu”.
Turning the Cal=Trak Off
The Cal=Trak has a battery saving automatic shut off system. After 65 minutes of inactivity, the Cal=Trak will shut off. Alternatively, the unit can be shut off manually by pressing the Reset button followed by the number 0 from the “Main Menu”.
LCD Backlight
The Cal=Trak’s LCD display includes a backlight function to illuminate the display. The default setting for the backlight is always on. If you wish to conserve battery power you may wish to turn the backlight off or enable the backlight only when inputting information on the keypad.
To turn the backlight off, from the “Main Menu” press the Light button one time. To enable the backlight only when inputting information on the keypad press the Light button a second time. To return to the default setting of always on, press the Light button a third time.
17