Sierra SL-800 instruction manual Setup Menu 2, Reading Type, # in Average & Minutes/Reading

Page 21

Option 3, “Cal. Type”

This option changes the sample type being displayed. Toggles between “std.” (standardized or mass flow readings as used by MFM or MFC devices) and “vol.” (volumetric readings as used on PD/rotameters)

8.7Setup Menu 2, Reading Type, # in Average & Minutes/Reading

Option 1, “Reading Type”

Toggles between “Cont” (continuous auto-read) and “Burst” (performs continuous auto-read, then stops after the quantity in averaging sequence is reached [as programmed below in Setup Menu 2, Option 2, # in Average])

Option 2, “# in Average”

Changes the quantity in an averaging sequence (consecutive readings) from 1 to 100. When used in “Burst” mode, the Cal=Trak stops after this number of readings.

Option 3, “Min./Reading”

Specifies the time interval (in minutes) between flow readings. A setting of “00” means continuous readings (no interval).

8.8Setup Menu 3, Pressure Units, Temp. Units, and Temp. Correction Factor

Option 1, “Pres. Units”

This option allows you to toggle between pressure units of: mmHg, kPa, mBar, or PSI

Option 2, “Temp. Units”

This option allows you to set the temperature units for ° F or ° C.

Option 3, “Temp Corr”

This option is used to set the standardization (mass flow) reference temperature. Use the keypad to enter the numerical value of the reference temperature. For mass flow units of slpm, a reference temperature of 21.1°C is commonly used. For mass flow units of nlpm, a reference temperature of 0.0°C is common. Check the specific reference conditions for each DUT.

8.9Setup Menu 4, Date, Time & Battery Voltage

This menu has an alternate navigation method.

21

Image 21
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page Theory of Operation General DescriptionIdealized Automatic Piston Prover Your Cal=Trak SL-800 Base Includes Cal=Trak Layout Unpacking ChecklistEach of Your Cal=Trak SL-800 Flow Cells Includes Connecting the Cal=Trak to a Flow Source Cal=Trak Installation Attaching & Removing Flow CellsCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Inventory Dead Volume Application PrecautionsComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Vacuum Comparison of Cal=Trak with Piston or Bell Provers Alternative Setup for Piston ProversComparison of Cal=Trak with Sonic Nozzle Transfer Standards Setup for LFEsCalibration of Mass Flow Controllers MFCs Vacuum Setup for Sonic Nozzle Transfer StandardCalibration of Rotameters Variable Area Flow Meters Calibration of Mass Flow Meters MFMsOperating Instructions Charging the Cal=Trak BatteryTurning the Cal=Trak On Turning the Cal=Trak Off Cal=Trak Keypad How To Use the Cal=Trak Keypad Low Battery Indicator General Menu Navigation Factory Default Settings Parameters Factory Settings Optional SettingsTaking Readings Option 1, Sensor Factor Setting User PreferencesSetup Menu 4, Date, Time & Battery Voltage Setup Menu 2, Reading Type, # in Average & Minutes/ReadingOption 1, Date Setup Menu 5, Date & Time FormatsOption 2, Time Option 3, Auto OffBattery System Battery Maintenance & StorageMaintenance To Initiate the Leak Test Quality Assurance Leak Test ProcedureLeak Test Illustration Leak Test in Progress CalibrationReplacement Parts & Accessories Returning Your Unit for Calibration or Service ShipmentCalibration Includes Part DescriptionAdditional Information Limited Warranty Settings have been lost or seem to have changed Charge LED does not light when charger pluggedPiston Error appears on display Unit suddenly turns offPiston does not move when the Read or Auto is pushed When taking readings, the flow seems offDuring Leak Test, Leakage 9.999 ccm appears on display Contact Sierra for Assistance Resetting the Cal=TrakFlow Cell Standardized 15-30C Model Optimum Flow RangeIntroduction Installation Operating Instructions for Cal=Soft SoftwareRequirements Worksheet descriptionsIntroduction Setup Cal=TrakSetup Cal=Soft Setup DUTPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Software Troubleshooting Serial Port Selection IssueSerial Port Configuration Issue