Sierra SL-800 instruction manual

Page 37

Reference temperature: Will be selected automatically based in the units selected (Normal= 0C, Standard=21.1C)

Reference pressure: Will be selected automatically based in the units selected

DUT inlet pressure: enter the inlet gas pressure to the DUT. (Used for certificate data only)

DUT outlet pressure: enter the outlet gas pressure of the DUT. (Used for certificate data only)

Calibration position: Enter the position in which your DUT is mounted during the calibration (Used for certificate data only)

Accuracy: Will be pulled down from the program’s instrument database "Instrument information" if you are calibrating a Sierra Model. If you are calibrating an instrument of another manufacturer, enter the rated accuracy of the DUT here.

Repeatability: same as Accuracy above.

Temperature coefficient: Will be pulled down from the database "Instrument information" if you are calibrating a Sierra Model. For other makes, enter the DUT coefficient here.

Pressure coefficient: Same as Temperature coefficient above.

Max Allowed deviation: Will be pulled down from the program’s database "Instrument information" based on the Sierra Model selected (This value is the pass/fail criteria as shown on the calibration certificate. It is in general either the accuracy of the meter + uncertainty of the calibrator or the RSS of the accuracy of both devices {Sq Root of [meter accy2 + calibrator accy2]}.

Step 3—Perform a calibration

Press the START button

This will begin the calibration routine. A red screen will appear in which several options are given to start the certification. Do nothing until the window: "Please wait" disappears. The screen looks like this:

Never select with your mouse the background sheet or another sheet until after you press "Exit" or the program could stop running.

37

Image 37
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page Theory of Operation General DescriptionIdealized Automatic Piston Prover Your Cal=Trak SL-800 Base Includes Cal=Trak Layout Unpacking ChecklistEach of Your Cal=Trak SL-800 Flow Cells Includes Connecting the Cal=Trak to a Flow Source Cal=Trak Installation Attaching & Removing Flow CellsCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Inventory Dead Volume Application PrecautionsComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Vacuum Comparison of Cal=Trak with Piston or Bell Provers Alternative Setup for Piston ProversComparison of Cal=Trak with Sonic Nozzle Transfer Standards Setup for LFEsCalibration of Mass Flow Controllers MFCs Vacuum Setup for Sonic Nozzle Transfer StandardCalibration of Rotameters Variable Area Flow Meters Calibration of Mass Flow Meters MFMsOperating Instructions Charging the Cal=Trak BatteryTurning the Cal=Trak On Turning the Cal=Trak OffLow Battery Indicator Cal=Trak Keypad How To Use the Cal=Trak KeypadGeneral Menu Navigation Parameters Factory Settings Optional Settings Factory Default SettingsTaking Readings Option 1, Sensor Factor Setting User PreferencesSetup Menu 4, Date, Time & Battery Voltage Setup Menu 2, Reading Type, # in Average & Minutes/ReadingOption 1, Date Setup Menu 5, Date & Time FormatsOption 2, Time Option 3, Auto OffBattery Maintenance & Storage Battery SystemMaintenance To Initiate the Leak Test Quality Assurance Leak Test ProcedureLeak Test Illustration Leak Test in Progress CalibrationReplacement Parts & Accessories Returning Your Unit for Calibration or Service ShipmentCalibration Includes Part DescriptionAdditional Information Limited Warranty Settings have been lost or seem to have changed Charge LED does not light when charger pluggedPiston Error appears on display Unit suddenly turns offWhen taking readings, the flow seems off Piston does not move when the Read or Auto is pushedDuring Leak Test, Leakage 9.999 ccm appears on display Contact Sierra for Assistance Resetting the Cal=TrakFlow Cell Standardized 15-30C Model Optimum Flow RangeIntroduction Installation Operating Instructions for Cal=Soft SoftwareRequirements Worksheet descriptions Introduction Setup Cal=TrakSetup Cal=Soft Setup DUTPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Serial Port Selection Issue Software TroubleshootingSerial Port Configuration Issue