Sierra SL-800 Cal=Trak Installation Attaching & Removing Flow Cells, Cal=Trak Measurement Cycle

Page 9

6.0Cal=Trak Installation

6.1Attaching & Removing Flow Cells

The Sierra Cal=Trak accepts interchangeable cells for different flow ranges. If user tries to enter “Run Menu” prior to installing a flow cell, the unit indicates “No Cell” and returns to the “Main Menu” after a 5 second delay.

Attaching Flow Cells

1.Position the selected flow cell into the electronics base opening with its silver Sierra Instruments label facing you.

2.Locate the guide pins; when the guide pins are engaged, press down firmly.

3.When the power is turned on, the base electronics will sense which cell is installed and display the appropriate units for that cell.

Removing Flow Cells

Grasp the flow cell firmly by the base of the cylinder and lift upward out of the base.

6.2Connecting the Cal=Trak to a Flow Source

As the accuracy of the Cal=Trak is dependent upon the mechanical set-up (plumbing) of the device under test, it is useful to review the basic operation of the calibrator prior to plumbing. Always remember the following important guidelines:

1.The accuracy of the Cal=Trak is dependent upon its source being stable. An unstable flow source may produce inconsistent readings.

2.Sierra Instruments’ Cal=Trak is designed to be used at ambient pressures. This is easily accomplished by leaving the outlet of the flow cell open to atmosphere for positive pressure installations or the inlet open to atmosphere for vacuum installations. If a vent hose is required on the outlet fitting, a maximum pressure of 5 inches water column (0.18 psi or 12 mbar) above ambient is acceptable. Exhaust pressure of more than this amount will add additional uncertainty to the flow measurements of Cal=Trak. One way to reduce exhaust pressure is to increase the diameter of the vent line. Do not subject the Cal=Trak to a differential pressure above 0.35 bar (5 PSI). In other words, the pressure drop across the Cal=Trak calibrator must not exceed 0.35 bar (5 PSI) or damage may occur. In vacuum scenarios, make certain that the pressure drawn across the Cal=Trak does not exceed 0.35 bar (5 PSI).

3.Flow direction is indicated by the arrow on the top of the flow cell. To use a pressure flow source, connect to the inlet fitting, or to use a vacuum flow source, connect to the outlet fitting.

6.3The Cal=Trak Measurement Cycle

Operation of a Cal=Trak is extraordinarily simple, and little training is required. However, any measurement interacts with the device being calibrated to some degree. Often, these interactions are negligible. However, sometimes device interactions can seriously affect measurement accuracy. Here we will explain what happens during a Cal=Trak measurement to aid in installing and using the instrument appropriately.

In its inactive state, the Cal=Trak will, like any device, exhibit a constant insertion pressure drop. At all but the highest flows, the pressure drop is very small. In the inactive state, gas flows from the inlet to the outlet through the bypass valve (Figure 3). When a measurement cycle begins, the bypass valve closes, and the gas is directed into the cylinder, effectively inserting the piston in series with the gas flow, allowing measurement. Timing commences after the piston has accelerated to the flow stream’s speed. At the end of the timed cycle, the valve opens and the piston falls to its inactive position at the bottom of the cylinder.

9

Image 9
Contents Cal=Trak SL-800 Primary Gas Flow Calibrator Worldwide Locations to Serve You Table of Contents Page Theory of Operation General DescriptionIdealized Automatic Piston Prover Your Cal=Trak SL-800 Base Includes Cal=Trak Layout Unpacking ChecklistEach of Your Cal=Trak SL-800 Flow Cells Includes Connecting the Cal=Trak to a Flow Source Cal=Trak Installation Attaching & Removing Flow CellsCal=Trak Measurement Cycle Attaching Flow CellsInitial Pressure Pulse Timing KPa Begins Ends Inventory Dead Volume Application PrecautionsComparison vs. Calibration Initial Pressure PulseSetup for Piston or Bell Provers Vacuum Comparison of Cal=Trak with Piston or Bell Provers Alternative Setup for Piston ProversComparison of Cal=Trak with Sonic Nozzle Transfer Standards Setup for LFEsCalibration of Mass Flow Controllers MFCs Vacuum Setup for Sonic Nozzle Transfer StandardCalibration of Rotameters Variable Area Flow Meters Calibration of Mass Flow Meters MFMsOperating Instructions Charging the Cal=Trak BatteryTurning the Cal=Trak On Turning the Cal=Trak OffCal=Trak Keypad How To Use the Cal=Trak Keypad Low Battery IndicatorGeneral Menu Navigation Factory Default Settings Parameters Factory Settings Optional SettingsTaking Readings Option 1, Sensor Factor Setting User PreferencesSetup Menu 4, Date, Time & Battery Voltage Setup Menu 2, Reading Type, # in Average & Minutes/ReadingOption 1, Date Setup Menu 5, Date & Time FormatsOption 2, Time Option 3, Auto OffBattery System Battery Maintenance & StorageMaintenance To Initiate the Leak Test Quality Assurance Leak Test ProcedureLeak Test Illustration Leak Test in Progress CalibrationReplacement Parts & Accessories Returning Your Unit for Calibration or Service ShipmentCalibration Includes Part DescriptionAdditional Information Limited Warranty Settings have been lost or seem to have changed Charge LED does not light when charger pluggedPiston Error appears on display Unit suddenly turns offPiston does not move when the Read or Auto is pushed When taking readings, the flow seems offDuring Leak Test, Leakage 9.999 ccm appears on display Contact Sierra for Assistance Resetting the Cal=TrakFlow Cell Standardized 15-30C Model Optimum Flow RangeIntroduction Installation Operating Instructions for Cal=Soft SoftwareRequirements Worksheet descriptionsIntroduction Setup Cal=TrakSetup Cal=Soft Setup DUTPage Page Automatic Flow Calibration Using Sierra 100 Series MFC Software Troubleshooting Serial Port Selection IssueSerial Port Configuration Issue