Teledyne OT-3 Trace Oxygen Analyzer Installation, Anatomy of a Micro-fuel Cell

Page 15

Trace Oxygen Analyzer

Installation

 

 

 

between a battery and a true fuel cell. (All of the reactants are stored externally in a true fuel cell.)

2.2.2 Anatomy of a Micro-fuel Cell

The Micro-fuel Cell is made of extremely inert plastic (which can be placed confidently in practically any environment or sample stream). It is effectively sealed, though one end is a gas permeable membrane. At the permeable end a screen retains a diffusion membrane through which the oxygen passes into the cell. At the other end of the cell is a printed circuit board with concentric gold contacts.

Refer to Figure 2-1,Cross Section of a Micro-fuel Cell, which illustrates the following internal description.

Figure 2-1: Cross Section of a Micro-fuel Cell

At the sensing end of the cell is a diffusion membrane, whose thickness is very accurately controlled. Near the diffusion membrane lies the oxygen sensing element—the cathode.

The anode structure is larger than the cathode. It is made of lead and is designed to maximize the amount of metal available for chemical reaction.

The space between the active elements is filled by a structure saturated with electrolyte. Cathode and anode are wet by this common pool. They each have a conductor connecting them, through some

Teledyne Analytical Instruments

7

Image 15
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Maintenance OperationAppendix Front Panel List of FiguresTrace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Micro-fuel Cell Sensor Installation OT-3Operational Theory IntroductionAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationInstalling the Micro-fuel Cell Location and MountingElectrical Connections Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation OT-3 LO Range Setting the Alarm SetpointsSet Alarm HI RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Maintenance Trace Oxygen Analyzer MaintenanceReplacing the Fuse Sensor Installation or Replacement Maintenance OT-3When to Replace a Sensor Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions Specifications AppendixAppendixOT-3 Spare Parts List Trace Oxygen AnalyzerAppendixAnalytical Instruments Reference Drawing MiscellaneousAppendix OT-3 Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments