Trace Oxygen Analyzer | Installation | |
|
|
|
The specific descriptions for each type of alarm are as follows:
Power Fail Alarm: This alarm changes state when the AC or DC system power is lost.
Alarm #1: Programmable as high or low alarm (actuates when concentration is above threshold). Can be set anywhere between 0 and 9999ppm, but must be set ABOVE the threshold set for Alarm 2.
Alarm #2: Programmable as high or low alarm (actuates when concentration is below threshold). Can be set anywhere from 0 to 9999ppm.
CAUTION: THERE COULD BE HAZARDOUS VOLTAGE AT THE ALARMS TERMINALS, EVEN WHEN POWER IS REMOVED FROM THE INSTRUMENT.
3.3.4 Relay Ratings
The
∙Higher rating
DC Rating:
AC Rating: 125 VAC maximum @ 0.5 Amps
minimum reliable switching rating: 10 μA @ 10mV
∙Lower
DC Rating:
AC Rating: 125 VAC maximum @ 0.2 Amps maximum
minimum reliable switching rating: 10 μA @ 10mV
To safely use the alarm relays in a hazardous location, the characteristics of both the switched voltage and the components or load being operated must be known. The alarm relays must not be attached to
Teledyne Analytical Instruments | 17 |