Teledyne OT-3 operating instructions Installing a Micro-fuel Cell

Page 42

Maintenance

OT-3

 

 

2.Open the analyzer door.

3.Unscrew the cell-retainer cap from the cell block by turning it counterclockwise until it is free.

4.Withdraw the cap from the block. The cell should come out with the cap.

5.Carefully pull the cell off of the cap.

WARNING: DO NOT TOUCH THE SCREENED END OF THE CELL OR ANY FLUID THAT MAY BE LEAKING FROM IT.

6.Dispose of the cell in a safe manner, in accordance with all applicable ENVIRONMENTAL AND SAFETY laws.

5.2.4Installing a Micro-fuel Cell

To install a new Micro-fuel Cell:

CAUTION: DO NOT SCRATCH, PUNCTURE, OR OTHERWISE DAMAGE THE SENSING MEMBRANE OF THE MICRO-FUEL CELL. IF THE MEMBRANE IS DAMAGED, THE CELL MUST BE REPLACED.

1.Examine the O-ring at the base of the cell-retainer cap, and replace it if it is worn of damaged.

2.Disconnect the Power Source from the Control Unit.

3.Set the flow rate to a high level (.5-1 SCFH).

4.As rapidly as possible complete steps 5-7.

5.Remove the new Micro-fuel Cell from its protective bag.

6.Place the cell on the end of cell-retainer cap with the concentric gold rings facing up.

7.Careful insert the cap and cell into the block, and screw the cap clockwise into the cell block until it is fully closed.

8.Reconnect the electrical power.

Teledyne Analytical Instruments

34

Image 42
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel OT-3 Copyright 2000 Teledyne Analytical InstrumentsSpecific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Maintenance OperationAppendix List of Figures Front PanelOverview Main Features of the AnalyzerIntroduction Trace Oxygen Analyzer IntroductionFront Panel Description Introduction OT-3Function Keys Rear Panel Description Measurement Parameters ReadoutRear Panel AC Version Introduction Installation OT-3Operational Theory Micro-fuel Cell SensorTrace Oxygen Analyzer Installation Anatomy of a Micro-fuel CellElectrochemical Reactions Calibration Characteristics Effect of PressureElectronics GeneralSignal Processing Block Diagram of the Signal Processing ElectronicsInstallation OT-3 Installation Unpacking the AnalyzerInstalling the Micro-fuel Cell Location and MountingElectrical Connections Primary Input Power Analog OutputsAlarm Relays Contact ID for Failsafe Relay OperationRelay Ratings Gas Connections Solid State Relay OutputGas Connections Sample System Installation Checklist OT-3 Sample SystemInstallation OT-3 Teledyne Analytical Instruments Operation Trace Oxygen Analyzer OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation OT-3 HI Range Setting the Alarm SetpointsSet Alarm LO RangeAlarm Delay Power Failure AlarmKeypad Lockout Selecting a Fixed Range or Autoranging CalibrationDisplaying Percent & PPM on the LED Display Supplementary InformationOperation OT-3 Maintenance Trace Oxygen Analyzer MaintenanceReplacing the Fuse Sensor Installation or Replacement Maintenance OT-3When to Replace a Sensor Ordering and Handling of Spare Sensors Removing the Micro-fuel CellInstalling a Micro-fuel Cell Cell Warranty Conditions Specifications AppendixAppendixOT-3 Spare Parts List Trace Oxygen AnalyzerAppendixAnalytical Instruments Reference Drawing MiscellaneousAppendix OT-3 Index Trace Oxygen Analyzer IndexIndex OT-3 No PassWebsite address Index OT-3 Teledyne Analytical Instruments