Teledyne OT-3 operating instructions Keypad Lockout

Page 35

Trace Oxygen Analyzer

Operation

 

 

 

When the alarms are set for delay and an alarm condition exists, the display will alternate between the O2 reading and the time in minutes and seconds before the alarm will trigger. The alarm indicator on the front panel will blink on and off. The delay time will continue to count down until it reaches zero then the alarm relay will close. At this point the display will stop blinking and the O2 concentration will display continuously. If a second alarm is reached after the delay time has elapsed, it will trigger immediately without any delay. If the O2 level drops to below the alarm condition during the delay, the delay timer will reset to the full delay.

If the alarm delay is set to 0 and an alarm condition exists, the alarm indicator will blink and the alarm relay will close immediately. When the O2 level drops out of the alarm condition, the alarm relay will drop out accordingly.

The alarm delay can be toggled on or off and the timing set through the DELAY function. To set the alarm delay:

1.Press the UP and DOWN arrow keys simultaneously. The display will indicate either PASS CODE or DELAY. If the display indicates PASS CODE, then to set the delay, press both arrow keys simultaneously again to toggle to DELAY.

2.While the display is blinking the word PASS DELAY, press the UP arrow key to increase the delay. To decrease the delay, press the DOWN arrow key. After three seconds the delay feature will time-out and return to the analyze mode.

4.6Keypad Lockout

The Model OT-3 has a keypad lockout feature which is designed to prevent unauthorized changes to the analyzer. This feature is toggled on or off via the PASS CODE function. To enable the keypad lockout:

1.Press the UP and DOWN arrow keys simultaneously. The display will indicate either PASS CODE or DELAY. If the display indicates DELAY, then to set the lockout feature, press both arrow keys simultaneously again to toggle to PASS CODE.

2.While the display is blinking the word PASS CODE, press the up or down arrow key to alter the code number. Any number except 10 will lock the keypad.

3.After three seconds, the control unit will time-out and return to analyze mode. With the keypad locked, pressing the keys

Teledyne Analytical Instruments

27

Image 35
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Front Panel List of FiguresTrace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Micro-fuel Cell Sensor Installation OT-3Operational Theory IntroductionAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationElectrical Connections Installing the Micro-fuel CellLocation and Mounting Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationOperation OT-3 Using the Function and Data Entry ButtonsSetting the Analysis Ranges LO Range Setting the Alarm SetpointsSet Alarm HI RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance When to Replace a Sensor Sensor Installation or ReplacementMaintenance OT-3 Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions AppendixOT-3 SpecificationsAppendix Analytical Instruments Spare Parts ListTrace Oxygen AnalyzerAppendix Appendix OT-3 Reference DrawingMiscellaneous Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments