Operation |
|
|
|
4.2Using the Function and Data Entry Buttons
When no buttons on the Analyzer are being pressed, the instrument is in the Analyze mode. It is monitoring the amount of oxygen in the sample gas that is flowing through the sampling system.
When one of the Function Buttons is being pressed, the Analyzer is in the Setup mode or the Calibration mode.
The 4 Setup function buttons on the analyzer are:
∙
∙
∙
∙
SET ALARM 1
SET ALARM 2
SET HI RANGE
SET LO RANGE
The Calibration mode button is:
∙SET SPAN
The Data Entry buttons (D and Ñ) increment the values displayed on the TRACE OXYGEN meter while one of the Function buttons is being held down.
D: UP
Ñ: DOWN
Any of the functions can be selected at any time by holding down the appropriate button.
Each function will be described in the following sections. Although the operator can use any function at any time, the order chosen in this manual is appropriate for an initial setup.
4.3Setting the Analysis Ranges
The two user definable analysis ranges are both capable of being adjusted for from
Note: The HI Range setpoint MUST be set at a higher concentration percentage than the LO Range setpoint.
Teledyne Analytical Instruments | 24 |