Teledyne OT-3 operating instructions Introduction, Overview, Main Features of the Analyzer

Page 9

Trace Oxygen Analyzer

Introduction

 

 

 

Introduction

1.1Overview

The Teledyne Electronic Technologies Analytical Instruments (TET/AI) Model OT-3 is a microprocessor-based trace oxygen analyzer for real-time measurement of the parts per million of oxygen in inert gases, or in a wide variety of gas mixtures. It features simple operation, fast response, and a compact, rugged construction. Typical applications of the Model OT-3 include monitoring natural gas pipelines, wellheads, and compressor/blowers.

1.2Main Features of the Analyzer

The main features of the analyzer include:

Insta-Trace sensor technology

Stainless steel sample system (with PVC scrubber and Nylon cell block)

Pressure regulator

Coalescing filter

Acid gas scrubber (clear, rebuildable )

Flow control restrictor

Sample shut off valve

Bypass/filter drain valve

Flowmeter

Viewing window

Sample/Span selector valve

NEMA 4 enclosure (15.8” x 11.8” x 8.1”).

Viewing window 3300/OT-3 Control Unit

Teledyne Analytical Instruments

1

Image 9
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Maintenance OperationAppendix Front Panel List of FiguresIntroduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Operational Theory Installation OT-3Introduction Micro-fuel Cell SensorAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationInstalling the Micro-fuel Cell Location and MountingElectrical Connections Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation OT-3 Set Alarm Setting the Alarm SetpointsHI Range LO RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Maintenance Trace Oxygen Analyzer MaintenanceReplacing the Fuse Sensor Installation or Replacement Maintenance OT-3When to Replace a Sensor Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions Specifications AppendixAppendixOT-3 Spare Parts List Trace Oxygen AnalyzerAppendixAnalytical Instruments Reference Drawing MiscellaneousAppendix OT-3 Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments