Teledyne OT-3 operating instructions Gas Connections

Page 27

Trace Oxygen Analyzer

Installation

 

 

 

Vent connection

Coalescing filter drain

Note: Zero gas is not required however, see Section 5.2.1.

Figure 3-3: Gas Connections

To connect the gas lines, remove the fitting and ferrule and cone at the bottom panel. Insert the tubing into the fitting then place the ferrule and cone over the tubing. Mate the assembly back to the connector on the panel and tighten finger tight. Use a wrench to tighten an additional

1 ¼ turn beyond finger tight. You may have to hold the stationary fitting with another wrench to keep the entire fitting from turning. Figure 3-3 shows the gas connection panel.

CAUTION: TIE IN ONE LINE AT A TIME. MAKE SURE THAT THE CORRECT GAS LINE IS MATED TO THE PROPER FITTING ON THE INSTRUMENT.

For maximum performance, the system should be installed in a manner to minimize the sample line length.

Teledyne Analytical Instruments

19

Image 27
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Maintenance OperationAppendix Front Panel List of FiguresTrace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction OverviewIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Micro-fuel Cell Sensor Installation OT-3Operational Theory IntroductionAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationInstalling the Micro-fuel Cell Location and MountingElectrical Connections Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationUsing the Function and Data Entry Buttons Setting the Analysis RangesOperation OT-3 LO Range Setting the Alarm SetpointsSet Alarm HI RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Maintenance Trace Oxygen Analyzer MaintenanceReplacing the Fuse Sensor Installation or Replacement Maintenance OT-3When to Replace a Sensor Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions Specifications AppendixAppendixOT-3 Spare Parts List Trace Oxygen AnalyzerAppendixAnalytical Instruments Reference Drawing MiscellaneousAppendix OT-3 Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments