Trace Oxygen Analyzer | Installation | |
|
|
|
∙
∙
Vent connection
Coalescing filter drain
Note: Zero gas is not required however, see Section 5.2.1.
Figure 3-3: Gas Connections
To connect the gas lines, remove the fitting and ferrule and cone at the bottom panel. Insert the tubing into the fitting then place the ferrule and cone over the tubing. Mate the assembly back to the connector on the panel and tighten finger tight. Use a wrench to tighten an additional
1 ¼ turn beyond finger tight. You may have to hold the stationary fitting with another wrench to keep the entire fitting from turning. Figure
CAUTION: TIE IN ONE LINE AT A TIME. MAKE SURE THAT THE CORRECT GAS LINE IS MATED TO THE PROPER FITTING ON THE INSTRUMENT.
For maximum performance, the system should be installed in a manner to minimize the sample line length.