Installation

 

 

 

A flowmeter and flow restrictor are parts of the sample system and assist in setting the flow of the gas. The restrictor is installed upstream of the analyzer while the flowmeter is installed on the vent side.

Adjust the sample pressure at the regulator to provide a flow rate of 0.4-0.5 SCFH. It is important that the sample vent connection does not restrict the sample flow. The sensor is designed to operate at atmospheric pressure. Restricting the sample vent line will pressurize the sensor, alter the O2 reading, and reduce the sample flow.

Sample and Span pressure should not exceed 2000 psig (13.8 Mpa)

Figure 3-4: OT-3 Sample System

3.6Installation Checklist

Before connecting the instrument to the power source and turning it on, make sure you have:

Correctly installed the Sample and Exhaust gas lines

Checked for leaks

Set the flow

Once the above checks have been made, you can connect to the power source. The instrument is now ready for operation.

Teledyne Analytical Instruments

21

Image 29
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 2000 Teledyne Analytical Instruments Model OT-3Specific Model Information Important Notice Trace Oxygen Analyzer Table of Contents Appendix MaintenanceOperation Front Panel List of FiguresIntroduction Main Features of the AnalyzerOverview Trace Oxygen Analyzer IntroductionIntroduction OT-3 Front Panel DescriptionFunction Keys Measurement Parameters Readout Rear Panel DescriptionRear Panel AC Version Operational Theory Installation OT-3Introduction Micro-fuel Cell SensorAnatomy of a Micro-fuel Cell Trace Oxygen Analyzer InstallationElectrochemical Reactions Effect of Pressure Calibration CharacteristicsGeneral ElectronicsBlock Diagram of the Signal Processing Electronics Signal ProcessingInstallation OT-3 Unpacking the Analyzer InstallationElectrical Connections Installing the Micro-fuel CellLocation and Mounting Analog Outputs Primary Input PowerContact ID for Failsafe Relay Operation Alarm RelaysRelay Ratings Solid State Relay Output Gas ConnectionsGas Connections Sample System OT-3 Sample System Installation ChecklistInstallation OT-3 Teledyne Analytical Instruments Trace Oxygen Analyzer Operation OperationOperation OT-3 Using the Function and Data Entry ButtonsSetting the Analysis Ranges Set Alarm Setting the Alarm SetpointsHI Range LO RangePower Failure Alarm Alarm DelayKeypad Lockout Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation OT-3 Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance When to Replace a Sensor Sensor Installation or ReplacementMaintenance OT-3 Removing the Micro-fuel Cell Ordering and Handling of Spare SensorsInstalling a Micro-fuel Cell Cell Warranty Conditions AppendixOT-3 SpecificationsAppendix Analytical Instruments Spare Parts ListTrace Oxygen AnalyzerAppendix Appendix OT-3 Reference DrawingMiscellaneous Trace Oxygen Analyzer Index IndexNo Pass Index OT-3Website address Index OT-3 Teledyne Analytical Instruments